Capillary Blood Glucose Monitor
Authors:
- 1 September 1989
- journal article
- research article
- Published by SAGE Publications in Clinical Pediatrics
- Vol. 28 (9) , 412-415
Abstract
The performance of a rapid capillary blood glucose monitor, the Glucoscan 2000, was evaluated in a neonatal patient population by comparison with a reference method on 283 paired patient samples. The median glucose was 56 mg/dL (3.1 mmol/L). The slope of the regression line was .93 with a correlation coefficient r of .82. The average error in Glucoscan values at low, intermediate and high glucose ranges was 43%, 20% and 12% respectively. Precision analysis by replicate determinations also showed an increased coefficient of variation, 7.4%, at low glucose values. The Glucoscan 2000 lacks sufficient accuracy and precision at low glucose values to serve as the exclusive analytic test for the diagnosis of neonatal hypoglycemia.Keywords
This publication has 12 references indexed in Scilit:
- Comparison of Five Glucose Meters for Self-Monitoring of Blood Glucose by Diabetic PatientsDiabetes Care, 1986
- Capillary Blood Glucose Measured By the Laboratory Phlebotomy TeamLaboratory Medicine, 1985
- Quality assessment of blood glucose monitors in use outside the hospital laboratory.Journal of Clinical Pathology, 1983
- Hypoglycaemia in infants and childrenClinics in Endocrinology and Metabolism, 1982
- Comparison of Methods for Blood Glucose MonitoringDiabetes Care, 1981
- Comparison of Various Methods for Rapid Glucose EstimationDiabetes Care, 1981
- Blood glucose monitors: a laboratory and patient assessment.BMJ, 1980
- Evaluation of a Direct-reading Reflectometer for Neonatal Hypoglycemia ScreeningAmerican Journal of Clinical Pathology, 1974
- Blood Glucose Determination in Infants and ChildrenAmerican Journal of Diseases of Children, 1972
- Neonatal hypoglycemia: A discussionThe Journal of Pediatrics, 1971