Negative interference from immunoglobulin M paraproteinaemia on the Roche enzymatic creatinine method
- 8 February 2022
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 59 (3) , 205-210
Abstract
Although enzymatic creatinine methods are subject to fewer interferences than traditional Jaffe creatinine methods, every method in clinical chemistry has limitations. We report, for the first time in the literature, a case of an immunoglobulin M (IgM) paraproteinaemia causing an undetectably low creatinine result on the Roche enzymatic assay. This interference did not occur with other enzymatic creatinine methods produced by Abbott and Siemens or the Roche Jaffe, VITROS dry slide and liquid chromatography with tandem mass spectrometry (LC-MS/MS) creatinine methods. IgM interference was confirmed as patient serum precipitated with polyethylene glycol (PEG) and anti-IgM antiserum yielded detectable Roche enzymatic creatinine results comparable to unaffected methods. The patient’s serum formed an obvious precipitate when mixed with reagent one of the Roche enzymatic creatinine method. This is in contrast to a report of positive interference from IgM paraproteinaemia in a different enzymatic creatinine method, which showed that a precipitate formed when mixing blood with reagent two. As each patient’s paraprotein has a unique structure, it is possible that there are variations in the chemical characteristics of IgM paraproteins between patients. This, as well as IgM-class antibodies’ tendency to form multimers and aggregates, can lead to unpredictable assay interferences and precipitation tendencies between different manufacturers of enzymatic creatinine reagents and their incubation steps. This case highlights the importance of continuing to question and investigates results that do not fit the clinical picture, especially as more laboratories switch from primarily using traditional Jaffe creatinine methods to enzymatic creatinine methods.Keywords
This publication has 17 references indexed in Scilit:
- Pseudohypercreatininaemia in two patients caused by monoclonal IgM interference with enzymatic assay of creatinineJournal of Clinical Pathology, 2015
- Phenindione interference in enzymatic creatinine assay – A case reportClinical Nephrology, 2015
- The applicability of eGFR equations to different populationsNature Reviews Nephrology, 2013
- NoticeKidney International Supplements, 2013
- Clinical evaluation of analytical variations in serum creatinine measurements: why laboratories should abandon Jaffe techniquesBMC Nephrology, 2012
- NoticeKidney International Supplements, 2012
- Specificity Characteristics of 7 Commercial Creatinine Measurement Procedures by Enzymatic and Jaffe Method PrinciplesClinical Chemistry, 2012
- How paraproteins can affect laboratory assays: spurious results and biological effectsPathology, 2010
- Pseudohypercreatininemia due to Positive Interference in Enzymatic Creatinine Measurements Caused by Monoclonal IgM in Patients with Waldenström’s MacroglobulinemiaNephron, 2000
- Analytical Reviews in Clinical Biochemistry: The Estimation of CreatinineAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1986