Sensitivity loss of Lif:Mg,Cu,P thermoluminescence dosemeters caused by oven annealing
Authors:
- F. GobletFGF. Goblet
- B. PolivkaBPB. Polivka
- H. SeifertHSH. Seifert
- 10 March 2006
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 121 (2) , 195-201
Abstract
LiF:Mg,Cu,P is a very sensitive thermoluminescence material that can be used for personal dosimetry in radiology. But if this material is heated too much during annealing or reading it quickly loses its sensitivity. This study shows that an annealing oven in wide use can cause thermal damage to the dosemeters owing to inhomogeneous temperature distribution in the annealing tray. At annealing temperatures >240°C, differences of only 1°C lead to significant losses of sensitivity. Therefore, it is necessary to measure the distribution of temperature in the annealing tray for correct placement of the dosemeters in the tray.Keywords
This publication has 8 references indexed in Scilit:
- OPTIMIZING THE USE OF LiF:MG,CU,P (GR-200 P) TO MEASURE LOW DOSE IRRADIATION IN NUCLEAR MEDICINEHealth Physics, 2003
- Handbook of Thermoluminescence2003
- Influence of Readout Parameters on TL Response, Re-usability and Residual Signal in LiF:Mg,Cu,PRadiation Protection Dosimetry, 2002
- Isothermal Decay Readout: Application to LiF:Mg,Cu,P and a-Al2O3:CRadiation Protection Dosimetry, 1999
- Evaluation of Bicron NE MCP DXT-RAD Passive Extremity DosemeterRadiation Protection Dosimetry, 1999
- Preparation and Characteristics of LiF:Mg,Cu,P Thermoluminescent MaterialRadiation Protection Dosimetry, 1993
- Influence of the Cooling Rate on Repeatability of LiF:Mg,Cu,P Thermoluminescent ChipsRadiation Protection Dosimetry, 1990
- Development of a new highly sensitive LiF thermoluminescence dosimeter and its applicationsNuclear Instruments and Methods, 1978