Variation in Mechanical Properties of 5.5 mm Cobalt Chromium and Titanium Spinal Rods Across Suppliers
- 25 May 2025
- journal article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 43 (8) , 1493-1500
Abstract
Cobalt chromium (CoCr) and titanium (Ti) alloys are popular spinal rod choices due to biocompatibility, less imaging artifact than stainless steel, and greater corrosion resistance. This study assesses the mechanical behavior in bending of commercially available 5.5 mm diameter CoCr and Ti spinal rods in a four‐point bending configuration. For each supplier, rod segments were loaded to 20 mm of deflection with continuous load and displacement recording. Rod mechanical properties were calculated. Suppliers were compared by ANOVA, with post hoc pairwise comparisons. For CoCr rods, flexural modulus and flexural stiffness varied across suppliers with maximum differences of 18.0 GPa and 47.0 N/mm, respectively, representing 11% and 15% variation. Range for CoCr rod yield load was 780–1004 N (29% variation), yield displacement was 2.3–2.8 mm, elastic displacement was 2.2–2.7 mm, and maximum load was 1791–2181 N, representing 22% variation. For Ti rods, flexural modulus and flexural stiffness varied across suppliers with maximum differences of 7.3 GPa and 24.3 N/mm, respectively, representing 9% and 14% variation. Range for Ti rod yield load was 1054–1207 N (14% variation), yield displacement was 5.9–7.0 mm, elastic displacement was 5.8–6.9 mm, and maximum load was 1515–1678 N, representing 11% variation. There is substantial variation in mechanical behavior in bending of CoCr and Ti rods (5.5 mm diameter) amongst suppliers, suggesting that they are not mechanically equivalent. Surgeons must ensure that the mechanical properties of a particular rod meet the requirements of the surgical correction required.Keywords
This publication has 18 references indexed in Scilit:
- Metals in SpineWorld Neurosurgery, 2017
- Comparative Mechanical Properties of Commonly Used Spinal RodsGlobal Spine Journal, 2015
- Influence of implant rod curvature on sagittal correction of scoliosis deformityThe Spine Journal, 2014
- Does the size of the rod affect the surgical results in adolescent idiopathic scoliosis? 5.5-mm versus 6.35-mm rodThe Spine Journal, 2014
- The Effect of Contouring on Fatigue Strength of Spinal Rods: Is it Okay to Re-bend and Which Materials Are Best?Spine Deformity, 2013
- Evaluation of spinal instrumentation rod bending characteristics for in‐situ contouringJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2011
- A comparison of magnetic and radiographic imaging artifact after using three types of metal rods: stainless steel, titanium, and vitalliumThe Spine Journal, 2010
- Corrosion of spinal implants retrieved from patients with scoliosisJournal of Orthopaedic Science, 2005
- Notch Sensitivity of Titanium Alloy, Commercially Pure Titanium, and Stainless Steel Spinal ImplantsSpine, 2001
- The Role of Spinal Instrumentation in Augmenting Lumbar Posterolateral FusionSpine, 1996