Scanning electron microscopy of the bone-bioactive implant interface
Authors:
- 15 September 1997
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 36 (4) , 429-440
Abstract
Rods of three bioactive materials, apatite/wollastonite glass ceramic (AW‐GC), bioactive glass (BG), and dense slip‐cast hydroxyapatite (HA), were implanted in the femora of 23 Wistar rats for periods of 1–4 weeks. The samples were harvested following vascular perfusion fixation and the femora freeze‐fractured for scanning electron microscopy to expose the bone/implant interface. The focus of our observations was when new bone was forming on the implant surfaces irrespective of the implantation period. Scanning microscopy of the hydroxyapatite rods demonstrated that in areas where bone bonding had occurred, the implant surface was composed of globular accretions which fused to form a cement‐like matrix to which collagen fibers were attached. Dissolution of individual grains of the implant surface created a roughened surface topography. Such features were not found in the transcortical portions of these implants. Similar globular accretions were also found on the surfaces of bulk AW‐GC, although bone apposition was not disrupted by the critical point‐drying procedure, and thus the interface was more difficult to image. Nevertheless, the collagen of the bony compartment interdigitated with an interfacial layer which was morphologically similar to that found on HA. The most surface reactive material, BG, demonstrated an interfacial structure where the surface reactive calcium phosphate layer was clearly distinguished from the underlying bulk implant material. However, this layer was separated from the overlying collagen‐containing bony compartment by a second, thinner, calcified layer which corresponded to the cement line matrix into which the collagen fibers were inserted. Our results show that the new bone interface formed with these three bioactive materials is morphologically comparable to that of cement lines found naturally in bone‐remodeling sites, and that this interfacial layer is formed on the chemically active surface of the biomaterial. The degree to which the cement line matrix interdigitated with the implant was a product of the reactivity of the implant surface. © 1997 John Wiley & Sons, Inc. J Biomed Mater Res, 36, 429–440, 1997.Keywords
This publication has 29 references indexed in Scilit:
- Initial bone matrix formation at the hydroxyapatite interface in vivoJournal of Biomedical Materials Research, 1995
- Ultrastructural study of the A‐W GC‐bone interface after long‐term implantation in rat and human boneJournal of Biomedical Materials Research, 1994
- Interface between hydroxyapatite and mandibular human bone tissueBiomaterials, 1992
- Ultrastructural differences of the interface zone between bone and Ti 6Al 4V or commercially pure titaniumJournal of Biomedical Engineering, 1989
- Interpretation of difficulties in the initial adhesion of bio-active glasses to boneBiomaterials, 1988
- The influence of bioglass on the growth of fibroblastsJournal of Biomedical Materials Research, 1987
- Formation of a high-strength bioactive glass-ceramic in the system MgO-CaO-SiO2-P2O5Journal of Materials Science, 1986
- Electron Microscopic Analysis of the Bone-Titanium InterfaceActa Orthopaedica, 1983
- Histochemical responses at a biomaterial's interfaceJournal of Biomedical Materials Research, 1974
- Bonding mechanisms at the interface of ceramic prosthetic materialsJournal of Biomedical Materials Research, 1971