Evaluation of an automatic threshold based detector of waveform limits in Holter ECG with the QT database
Authors:
- R. JaneR. Jane
- A. BlasiABA. Blasi
- J. GarciaJGJ. Garcia
- P. LagunaP. Laguna
- 22 November 2002
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 02766547,p. 295-298
Abstract
The authors evaluate a single-lead threshold based ECG wave boundaries detector with a QT database developed for validation purposes. They also identify its different sources of error distinguishing those that come from precision errors in boundary location from those that come from morphology misclassification. They obtain 71 % of records with correct morphology identification of T wave and variance in boundary location within manual referees variance. The remaining records analyzed correspond to signals with poor SNR at the T wave, or morphology discrepancies between algorithm and experts.Keywords
This publication has 4 references indexed in Scilit:
- Testing results and derivation of minimum performance criteria for computerized ECG-analysis2002
- A database for evaluation of algorithms for measurement of QT and other waveform intervals in the ECG2002
- Automatic Detection of Wave Boundaries in Multilead ECG Signals: Validation with the CSE DatabaseComputers and Biomedical Research, 1994
- New algorithm for QT interval analysis in 24-hour Holter ECG: performance and applicationsMedical & Biological Engineering & Computing, 1990