A 500 Hz predictive visual servoing scheme to mechanically filter complex repetitive organ motions in robotized surgery
Authors:
- L. SolerLSL. Soler
- J. LeroyJLJ. Leroy
- J. MarescauxJMJ. Marescaux
- 23 April 2004
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 3, 3361-3366
Abstract
Periodic deformations of organs and soft tissues are complex, repetitive disturbances for surgeons manipulating robotic interfaces in computer-assisted surgery. They are due to respiratory movements or heart beats, and they have to be manually compensated for by the surgeon whenever accurate gestures are needed, as it is the case in cardiac or robotized laparoscopic surgery. This work presents a repetitive model predictive control scheme for the cancellation of fast periodic motions by a robot arm, which is controlled by visual servoing at 500 Hz by means of a high-speed camera. The problem we address is to keep a constant distance in the camera images from a surgical tool's tip to the organ surface. Contributions of the control input to reference tracking and to the fast-disturbance rejection are split and computed separately to ensure that the surgeon's interaction on the robot bas no influence on the cancellation performance. The system is tested in a laboratory experiment with an experimental surgical arm and in in vivo conditions on a living pig with a standard surgical robot. Results show the effectiveness and the potential of the proposed control scheme.Keywords
This publication has 11 references indexed in Scilit:
- Combined image-based and depth visual servoing applied to robotized laparoscopic surgery2003
- Model Predictive Control for Cancellation of Repetitive Organ Motions in Robotized Laparoscopic SurgeryLecture Notes in Computer Science, 2003
- Heartbeat synchronization for robotic cardiac surgery2002
- Transatlantic robot-assisted telesurgeryNature, 2001
- Model predictive control of continuous yeast bioreactors using cell population balance modelsChemical Engineering Science, 2000
- Repetitive model predictive control applied to a simulated moving bed chromatography systemComputers & Chemical Engineering, 2000
- Robotic Motion Compensation for Respiratory Movement during RadiosurgeryComputer Aided Surgery, 2000
- Model Predictive Control1999
- Precise laser incisions, corrected for patient respiration with an intelligent aiming systemLasers in Surgery and Medicine, 1997
- A tutorial on visual servo controlIEEE Transactions on Robotics and Automation, 1996