Abstract
Children receive high radiation doses during interventional cardiology procedures and their sensitivity to radiation is higher than that of adults. A 4-year follow-up study was conducted in two large cardiovascular hospitals and a pediatric polyclinic with the objective of providing typical values. The median values of 454 cases obtained for fluoroscopy time, cumulative reference point air kerma (Ka,r) and air kerma-area product (PKA) by age range were 5.6 min, 53.6 mGy, and 4.1 Gy cm2 for <1 y; 4.3 min, 77.9 mGy, and 7.2 Gy cm2 for 1 to <5 y; 4.0 min, 41.4 mGy, and 7.4 Gy cm2 for 5 to <10 y; 4.4 min, 90.1 mGy, and 16.1 Gy cm2 for 10 to <16 y, respectively. The median values of 144 cases obtained for Ka,r and PKA by weight bands were 84.4 mGy and 7.4 Gy cm2 for 50 kg. Our results can serve as a dose comparison with local practice and as a basis for optimization between facilities.
Funding Information
  • Capital’s Funds for Health Improvement and Research, CFH (2022-2G-3019)
  • Capital’s Funds for Health Improvement and Research, CFH (2024-2G-30119)

This publication has 16 references indexed in Scilit: