Comparison of conventional and torsional mode (OZil) phacoemulsification: randomized prospective clinical study
Authors:
Open Access
- 1 November 2009
- journal article
- research article
- Published by SAGE Publications in European Journal of Ophthalmology
- Vol. 19 (6) , 984-989
Abstract
Purpose: To compare the amount of intraoperative energy used during conventional and torsional phacoemulsification surgery and investigate the changes in corneal endothelial cell density. Methods: In this study, a total of 100 eyes in patients with age-related cataract (mean age 68.9 years) were included. Patients were randomly allocated into conventional (group 1) or torsional (OZil, group 2) phacoemulsification. Mean nuclear grade made according to Lens Opacities Classification System II was not different in the 2 groups. Total ultrasound time (USTT), cumulative dissipated energy (CDE), and the percent of total equivalent power in position 3 (%USTEPiP3) were compared between the 2 groups. Results: There were 47 eyes in group 1 and 53 eyes in group 2. Patient age and sex were matched between the 2 groups (p=0.49, p=0.08). Mean USTT was 1.6±1.1 minutes (m) in group 1 and 1.5 ± 0.9 m in group 2 (p=0.55). Average CDE was 29.9±16.9 in group 1 and 25.2±19.1 in group 2 (p=0.20). Mean %USTEPiP3 was 27.0%±8.5% in group 1; however, it was 22.4%±6.6% in group 2 (p=0.003). At the first postoperative day, mean corneal edema was graded as 0.25±0.5 in group 1 and 0.18 ± 0.4 in group 2 (p=0.47). At the first postoperative week, it was found to be 0.02±0.15 in group 1 and 0.03±0.27 in group 2 (p=0.71). ECD loss was 6.7%±3.3% in group 1 and 4.2%±5.7% in group 2 (p=0.56). Conclusions: No statistically significant difference was found between conventional and torsional phacoemulsification with respect to USTT and CDE. Torsional mode led to a significant reduction in US power in footswitch position 3. Torsional mode also appears to be associated with less loss of ECDKeywords
This publication has 18 references indexed in Scilit:
- Torsional mode versus conventional ultrasound mode phacoemulsificationJournal of Cataract & Refractive Surgery, 2007
- Ultrasound energy and endothelial cell loss with stop-and-chop and nuclear preslice phacoemulsificationJournal of Cataract & Refractive Surgery, 2006
- Corneal endothelial damage after cataract surgery: Microincision versus standard techniqueJournal of Cataract & Refractive Surgery, 2006
- Thermography used for analysis and comparison of different cataract surgery procedures based on phacoemulsificationPhysiological Measurement, 2006
- Corneal endothelial function after phacoemulsification using the fluid-based system compared to conventional ultrasound techniqueEye, 2006
- A Survey About Phacoemulsification Incision Thermal Contraction Incidence and Causal RelationshipsAmerican Journal of Ophthalmology, 2006
- Corneal Endothelium Evaluation After Phacoemulsification With Continuous Anterior Chamber InfusionCornea, 2005
- Simultaneous surgeon and side-view video analysis comparing in situ fracture and stop-and-chop phacoemulsificationJournal of Cataract & Refractive Surgery, 2005
- Confocal microscopic observations of the human cornea following overnight contact lens wearClinical and Experimental Optometry, 2002
- Phacoemulsification and thermal wound injurySeminars in Ophthalmology, 2002