Quantitative accuracy in total-body imaging using the uEXPLORER PET/CT scanner
- 11 October 2021
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 66 (20) , 205008
Abstract
Absolute quantification of regional tissue concentration of radioactivity in positron emission tomography (PET) is a critical parameter-of-interest across various clinical and research applications and is affected by a complex interplay of factors including scanner calibration, data corrections, and image reconstruction. The emergence of long axial field-of-view (FOV) PET systems widens the dynamic range accessible to PET and creates new opportunities in reducing scan time and radiation dose, delayed or low radioactivity imaging, as well as kinetic modeling of the entire human. However, these imaging regimes impose challenging conditions for accurate quantification due to constraints from image reconstruction, low count conditions, as well as large and rapidly changing radioactivity distribution across a large axial FOV. We comprehensively evaluated the quantitative accuracy of the uEXPLORER total-body scanner in conditions that encompass existing and potential imaging applications (such as dynamic imaging and ultralow-dose imaging) using a set of total-body specific phantom and human measurements. Through these evaluations we demonstrated a relative count rate accuracy of ±3%–4% using the NEMA NU 2-2018 protocol, an axial uniformity spread of ±3% across the central 90% axial FOV, and a 3% activity bias spread from 17 to 474 MBq 18 F-FDG in a 210 cm long cylindrical phantom. Region-of-interest quantification spread of 1% was found by simultaneously scanning three NEMA NU 2 image quality phantoms, as well as relatively stable volume-of-interest quantification across 0.2%–100% of total counts through re-sampled datasets. In addition, an activity bias spread of −2% to +1% post-bolus injections in human subjects was found. Larger bias changes during the bolus injection phase in humans indicated the difficulty in providing accurate PET data corrections for complex activity distributions across a large dynamic range. Our results overall indicated that the quantitative performance achieved with the uEXPLORER scanner was uniform across the axial FOV and provided the accuracy necessary to support a wide range of imaging applications.Keywords
Funding Information
- National Institutes of Health (R01 CA206187)
- United Imaging Healthcare (PI research grant for Ramsey D Badawi and Simon R Cherry)
This publication has 28 references indexed in Scilit:
- Corrigendum: Optimal whole-body PET scanner configurations for different volumes of LSO scintillator: a simulation studyPhysics in Medicine & Biology, 2012
- Optimal whole-body PET scanner configurations for different volumes of LSO scintillator: a simulation studyPhysics in Medicine & Biology, 2012
- Physical and clinical performance of the mCT time-of-flight PET/CT scannerPhysics in Medicine & Biology, 2011
- Bias in iterative reconstruction of low-statistics PET data: benefits of a resolution modelPhysics in Medicine & Biology, 2011
- Quantifying and Reducing the Effect of Calibration Error on Variability of PET/CT Standardized Uptake Value MeasurementsJournal of Nuclear Medicine, 2011
- Reproducibility and Accuracy of Quantitative Myocardial Blood Flow Assessment with 82Rb PET: Comparison with 13N-Ammonia PETJournal of Nuclear Medicine, 2009
- From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid TumorsJournal of Nuclear Medicine, 2009
- Standards for PET Image Acquisition and Quantitative Data AnalysisJournal of Nuclear Medicine, 2009
- Accuracy of 3-Dimensional Reconstruction Algorithms for the High-Resolution Research TomographJournal of Nuclear Medicine, 2008
- Performance evaluation of the new whole-body PET/CT scanner: Discovery STEuropean Journal of Nuclear Medicine and Molecular Imaging, 2004