Investigation of uncertainties in image registration of cone beam CT to CT on an image-guided radiotherapy system
Authors:
- J R SykesJSJ R Sykes
- D S BrettleDBD S Brettle
- D R MageeDMD R Magee
- D I ThwaitesDTD I Thwaites
- 20 November 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (24) , 7263-7283
Abstract
Methods of measuring uncertainties in rigid body image registration of fan beam computed tomography (FBCT) to cone beam CT (CBCT) have been developed for automatic image registration algorithms in a commercial image guidance system (Synergy, Elekta, UK). The relationships between image registration uncertainty and both imaging dose and image resolution have been investigated with an anthropomorphic skull phantom and further measurements performed with patient images of the head. A new metric of target registration error is proposed. The metric calculates the mean distance traversed by a set of equi-spaced points on the surface of a 5 cm sphere, centred at the isocentre when transformed by the residual error of registration. Studies aimed at giving practical guidance on the use of the Synergy automated image registration, including choice of algorithm and use of the Clipbox are reported. The chamfer-matching algorithm was found to be highly robust to the increased noise induced by low-dose acquisitions. This would allow the imaging dose to be reduced from the current clinical norm of 2 mGy to 0.2 mGy without a clinically significant loss of accuracy. A study of the effect of FBCT slice thickness/spacing and CBCT voxel size showed that 2.5 mm and 1 mm, respectively, gave acceptable image registration performance. Registration failures were highly infrequent if the misalignment was typical of normal clinical set-up errors and these were easily identified. The standard deviation of translational registration errors, measured with patient images, was 0.5 mm on the surface of a 5 cm sphere centred on the treatment centre. The chamfer algorithm is suitable for routine clinical use with minimal need for close inspection of image misalignment.Keywords
This publication has 27 references indexed in Scilit:
- Quality Assurance for the Geometric Accuracy of Cone-Beam CT Guidance in Radiation TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Online adaptive radiotherapy of the bladder: Small bowel irradiated-volume reductionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- A deformable image registration method to handle distended rectums in prostate cancer radiotherapyMedical Physics, 2006
- Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Prostate gland motion assessed with cine-magnetic resonance imaging (cine-MRI)International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Accurate technique for complete geometric calibration of cone‐beam computed tomography systemsMedical Physics, 2005
- Medical image registrationPhysics in Medicine & Biology, 2001
- A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targetsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1999
- Visual assessment of the accuracy of retrospective registration of MR and CT images of the brainIEEE Transactions on Medical Imaging, 1998
- Determination of a Mean Orientation from a Cloud of Orientations. Application to Electron Back-Scattering Pattern MeasurementsJournal of Applied Crystallography, 1996