Microfabricated sol-gel relative humidity sensors for soil suction measurement during laboratory tests
- 1 August 2017
- journal article
- Published by Canadian Science Publishing in Canadian Geotechnical Journal
- Vol. 54 (8) , 1176-1183
Abstract
Currently there are no small sensors that can be incorporated inside soil samples for laboratory testing, to monitor water transport during loading. This is an important limitation to a better understanding of the hydromechanical coupled behaviour of soils. A sol-gel relative humidity sensor (11 mm × 11 mm), microfabricated in a clean room environment, was conceived to be incorporated in soil specimens during standard laboratory tests. The sensor operates based on changes in electrical resistivity detected by a cerium-doped silica–titania film deposited using a sol-gel technique over interdigitated aluminium electrodes spaced at 300 μm. To the best of the authors’ knowledge, sol-gel sensors for relative humidity measurement have never been used in soils; therefore, this is a novel application. The water retention curve of compacted kaolin was measured with the sensors and compared with the curve found using water dewpoint potentiometer WP4-C. The sensors were also tested incorporated in an oedometer cell, in which load was applied under vapour equilibrium. It was possible to detect the increment of the degree of saturation during compression. The use of the developed sensors incorporated in soils is considered acceptable for suction ranges between 1 and 10 MPa, which extends the suction interval covered by tensiometers, normally operating up to 2 MPa. Although the sensors require improvements in terms of sol-gel deposition and calibration protocol, the results confirm their scientific potential for being used in testing and characterization of unsaturated soils.Keywords
This publication has 19 references indexed in Scilit:
- Humidity sensing properties of mesoporous iron oxide/silica composite prepared via hydrothermal processSensors and Actuators B: Chemical, 2011
- Improved Calibration Functions of Three Capacitance Probes for the Measurement of Soil Moisture in Tropical SoilsSensors, 2011
- Benchmark of experimental techniques for measuring and controlling suctionGéotechnique, 2011
- Frequency, electrical conductivity and temperature analysis of a low-cost capacitance soil moisture sensorJournal of Hydrology, 2008
- Total suction measurement of unsaturated soils with a device using the chilled-mirror dew-point techniqueGéotechnique, 2003
- Polymer Capacitance Sensors for Measuring Soil Gas Humidity in Drier SoilsGeotechnical Testing Journal, 2003
- Calibration of tensiometer for direct measurement of matric suctionGéotechnique, 2003
- The influence of soil structure and stress history on the soil–water characteristics of a compacted tillGéotechnique, 1999
- Ceramic sensors for humidity detection: the state-of-the-art and future developmentsSensors and Actuators B: Chemical, 1995
- A Closed‐form Equation for Predicting the Hydraulic Conductivity of Unsaturated SoilsSoil Science Society of America Journal, 1980