Magnetic Resonance Visibility, Artifacts, and Overall Safety of the Self-Locating Peritoneal Dialysis Catheter with a Tungsten Tip
Open Access
- 24 January 2023
- journal article
- research article
- Published by Hindawi Limited in International Journal of Nephrology
- Vol. 2023, 1-6
Abstract
Background. The self-locating peritoneal dialysis (PD) catheter, contains a tungsten tip. The effects of magnetic resonance (MR) on the catheter were evaluated, emphasizing its MR signal, artifacts, ferromagnetism, and possible heating production during the MR sequences. Methods. The catheter was studied in an ex vivo model using a 1.5TMR system and placed into a plastic box containing saline solution. Acquisitions on coronal and axial planes were obtained on fast gradient-echo T1-weighted and fast spin-echo T2-weighted. In vivo abdominal MR exams were also carried out. Results. Overall, the catheter had good visibility. In all sequences, an extensive paramagnetic blooming artifact was detected at the level of the tip tungsten ballast, with a circular artifact of 5cm in diameter. The catheter showed no magnetic deflection, rotation, or movements during all MR sequences. After imaging, the temperature of the saline solution did not change compared to the basal measurement. Patients safely underwent abdominal MR. Conclusions. The results point to the possibility of safely performing MR in PD patients carrying the self-locating catheter. The self-locating PD catheter is stable when subjected to a 1.5TMR system. However, it creates some visual interference, preventing an accurate study of the tissues surrounding the tungsten tip.Keywords
This publication has 21 references indexed in Scilit:
- Safety of retained microcatheters: an evaluation of radiofrequency heating in endovascular microcatheters with nitinol, tungsten, and polyetheretherketone braiding at 1.5 T and 3 TJournal of NeuroInterventional Surgery, 2013
- ACR Guidance Document for Safe MR Practices: 2007American Journal of Roentgenology, 2007
- RF artifacts caused by metallic implants or instruments which get more prominent at 3 T: an in vitro studyMagnetic Resonance Imaging, 2005
- MR peritoneography in complications of continuous ambulatory peritoneal dialysisAbdominal Radiology, 2005
- American College of Radiology White Paper on MR SafetyAmerican Journal of Roentgenology, 2002
- Complications of Continuous Ambulatory Peritoneal DialysisAmerican Journal of Roentgenology, 2000
- Magnetic Materials and Properties2000
- MRI compatibility and visibility assessment of implantable medical devicesJournal of Magnetic Resonance Imaging, 1999
- Artifacts in magnetic resonance imaging from metalsJournal of Applied Physics, 1996
- Susceptibility artefacts in NMR imagingMagnetic Resonance Imaging, 1985