Single Droplet Separations and Surface Partition Coefficient Measurements Using Laser Ablation Mass Spectrometry
Authors:
- 3 November 2009
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 81 (23) , 9682-9688
Abstract
Surface activity of analytes plays a significant role in many chemical and physical phenomena. We present here a mass spectrometric method to characterize surface activity and solute partitioning between bulk liquid and the gas−liquid interface in droplets. The approach employs ablation by an infrared (IR) laser from the surface of a microliter droplet deposited on a stainless steel post. The ablated material is ionized for mass spectrometric analysis by either droplet charging or by postionization in an electrospray plume. Three areas of application have been explored using this method (1) separations in a single droplet: continuous ablation by a series of many successive laser pulses results in faster depletion of more surface active analytes, effectively comprising a surface activity-based separation. (2) Partition coefficient measurements: droplet volume is held constant during ablation by continually replenishing lost solvent. This leads to analyte-specific ion signal decay curves that may be fitted to a model based on Langmuir adsorption isotherms and simple analytical expressions, yielding quantitative values for the analyte surface partition coefficients. (3) Studies of the relationship between surface partitioning and high-performance liquid chromatography (HPLC) phase partitioning: comparisons of surface activities measured by laser desorption with retention times in reversed-phase HPLC reveal that the relationship between the two partitioning processes is very sensitive to chemical structure. Poor correlation between the retention time and surface activity is also observed within a subcategory of analytes (peptides). This effect is attributed to multimodal solute−stationary phase interactions. The laser desorption approach presented here provides direct information on analyte surface activities free from the complications encountered in chromatographic methods due to chemical structure variations.Keywords
This publication has 22 references indexed in Scilit:
- Atmospheric pressure infrared (10.6 µm) laser desorption electrospray ionization (IR‐LDESI) coupled to a LTQ Fourier transform ion cyclotron resonance mass spectrometerRapid Communications in Mass Spectrometry, 2009
- Development and Characterization of an Ionization Technique for Analysis of Biological Macromolecules: Liquid Matrix-Assisted Laser Desorption Electrospray IonizationAnalytical Chemistry, 2008
- Charge Assisted Laser Desorption/Ionization Mass Spectrometry of DropletsJournal of the American Society for Mass Spectrometry, 2008
- Mass Spectrometry of Acoustically Levitated DropletsAnalytical Chemistry, 2008
- Experimental Anion Affinities for the Air/Water InterfaceThe Journal of Physical Chemistry B, 2006
- On the mechanism of ion formation from the aqueous solutions irradiated with 3 µm IR laser pulses under atmospheric pressureJournal of Mass Spectrometry, 2006
- ON THE NATURE OF IONS AT THE LIQUID WATER SURFACEAnnual Review of Physical Chemistry, 2006
- Probing Interfacial Chemistry of Single Droplets with Field-Induced Droplet Ionization Mass Spectrometry: Physical Adsorption of Polycyclic Aromatic Hydrocarbons and Ozonolysis of Oleic Acid and Related CompoundsAnalytical Chemistry, 2006
- Ultra‐low flow nanospray for the normalization of conventional liquid chromatography/mass spectrometry through equimolar response: standard‐free quantitative estimation of metabolite levels in drug discoveryRapid Communications in Mass Spectrometry, 2006
- Adsorption Behavior and Dilational Rheology of the Cationic Alkyl Trimethylammonium Bromides at the Water/Air InterfaceThe Journal of Physical Chemistry B, 2005