Continuous Monitoring of Cardiac Rhythms in Left Ventricular Assist Device Patients
- 1 August 2013
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 38 (3) , 191-198
Abstract
Monitoring of cardiac rhythms is of major importance in the treatment of heart failure patients with left ventricular assist devices (LVADs) implanted. A continuous surveillance of these rhythms could improve out‐of‐hospital care in these patients. The aim of this study was to investigate cardiac rhythms using available pump data only. Datasets (n = 141) obtained in the normal ward, in the intensive care unit, and during bicycle ergometry were analyzed in 11 recipients of a continuous flow LVAD (59.1 ± 9.7 years; male 82%). Tachograms and arrhythmic patterns derived from the pump flow waveform, and a simultaneously recorded ECG were compared, as well as heart rate variability parameters such as: the average heart beat duration (RR interval), the standard deviation of the beat duration (SDNN), the root‐mean‐square of the difference of successive beat durations (RMSSD), and the number of pairs of adjacent beat duration differing by >50 ms divided by the number of all beats (pNN50). A very good agreement of cardiac rhythm parameters from the pump flow compared with ECG was found. Tachycardia, atrial fibrillation, and extrasystoles could be accurately identified from the tachograms derived from the pump flow. Also, Bland–Altman analysis comparing pump flow with ECG indicated a very small difference in average RR interval of 0.3 ± 1.0 ms, in SSDN of 0.5 ± 2.7 ms, in RMSSD of 1.0 ± 5.6 ms, and in pNN50 of 0.3 ± 1.0%. Continuous monitoring of cardiac rhythms from available pump data is possible. It has the potential to reduce the out‐of‐hospital diagnostic burden and to permit a more efficient adjustment of the level of mechanical support.Keywords
This publication has 23 references indexed in Scilit:
- Fifth INTERMACS annual report: Risk factor analysis from more than 6,000 mechanical circulatory support patientsThe Journal of Heart and Lung Transplantation, 2013
- Recommendations for the Use of Mechanical Circulatory Support: Device Strategies and Patient SelectionCirculation, 2012
- Arrhythmias in patients with ventricular assist devicesCurrent Opinion in Cardiology, 2012
- Evaluation of Left Ventricular Relaxation in Rotary Blood Pump Recipients Using the Pump Flow Waveform: A Simulation StudyArtificial Organs, 2011
- Left Ventricular Assist Improves Autonomic Imbalance in Patients with Persistent Myocardial DysfunctionThe International Journal of Artificial Organs, 2010
- Continuous assessment of cardiac function during rotary blood pump support: A contractility index derived from pump flowThe Journal of Heart and Lung Transplantation, 2010
- Heart rate variability in myocardial infarction and heart failureInternational Journal of Cardiology, 2007
- Atrial fibrillation in heart failure: epidemiology, pathophysiology, and rationale for therapyThe American Journal of Cardiology, 2003
- Impact of Reduced Heart Rate Variability on Risk for Cardiac EventsCirculation, 1996
- Heart Rate VariabilityCirculation, 1996