Integration of Low-Power Microfluidic Pumps with Biosensors within a Laboratory-on-a-Chip Device
Authors:
- 14 January 2009
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 81 (4) , 1365-1370
Abstract
We describe the fabrication of a controllable microfluidic valve coupled with an electrochemical pump, which has been designed to deliver reagents to an integrated microfluidic biosensing system. Fluid, retained within an insertion reservoir using a stop valve, was pumped using electrochemical actuation, providing a low power, low voltage integrated Laboratory-on-a-Chip for reproducible, small volume fluidic manipulation. The properties of the valve were characterized using both X-ray photoelectron spectroscopy and contact angle measurements, enabling the calculation of the magnitude of the forces involved (which were subsequently verified through experimental measurement). Electrochemical generation of oxygen and hydrogen acted as an on-demand pressure system to force fluid over the stop valve barrier. The process of filling-up the biosensing chamber was characterized in terms of the time to fill, the energy used, and the peak power consumed. The potential of the device was illustrated using a glucose biosensor.Keywords
This publication has 23 references indexed in Scilit:
- A refillable microfabricated drug delivery device for treatment of ocular diseasesLab on a Chip, 2008
- Surface-Enhanced Raman Signatures of Pigmentation of Cyanobacteria from within Geological Samples in a Spectroscopic-Microfluidic Flow CellAnalytical Chemistry, 2007
- Implementation of radiotelemetry in a lab-in-a-pill formatLab on a Chip, 2006
- An Electrochemical Pumping System for On-Chip Gradient GenerationAnalytical Chemistry, 2004
- Phase Change Microvalve for Integrated DevicesAnalytical Chemistry, 2004
- Quantitatively controlled nanoliter liquid manipulation using hydrophobic valving and control of surface wettabilityJournal of Micromechanics and Microengineering, 2002
- Nanofluidic Bubble Pump Using Surface Tension Directed Gas InjectionAnalytical Chemistry, 2002
- A closed-loop controlled electrochemically actuated micro-dosing systemJournal of Micromechanics and Microengineering, 2000
- Nanoliter Liquid Metering in Microchannels Using Hydrophobic PatternsAnalytical Chemistry, 2000
- An electrochemical microactuator: principle and first resultsJournal of Microelectromechanical Systems, 1996