Relationship Between Implant Connection and Implant Fracture: Systematic Review
Open Access
- 23 March 2025
- journal article
- review article
- Published by MDPI AG in Bioengineering
- Vol. 12 (4) , 333
Abstract
The design of dental implants has undergone minor modifications over the years to reduce possible future complications that may arise from implant rehabilitation. One critical element from a clinical point of view is the implant connection. Given this, the team of authors of the present study decided to biomechanically analyze the effect of implant connection obtained in a possible implant fracture through a systematic review of the published literature. To this end, a search was conducted in the PubMed and Scopus databases. Only finite element studies carried out in vitro and simulation studies were included, discarding clinical studies and related reviews. A total of 19 studies were obtained for analysis and complete study. The conical type is the connection design that demonstrates the best biomechanical behavior. However, there was also significant heterogeneity in the design of the studies, which translates into a substantial source of bias, as well as different types of specific designs within the conical connection. In conclusion, it was established that the design of the connection seems to play a fundamental role in the fatigue resistance of the implant when subjected to load. However, more studies are needed to determine the most optimal specific design.Keywords
This publication has 43 references indexed in Scilit:
- The impact of biocorrosion and titanium ions release on peri-implantitisClinical Oral Investigations, 2025
- Screening dental implant design parameters for effect on the fatigue limit of reduced-diameter implantsDental Materials, 2025
- Relationship between dental implant macro-design and osseointegration: a systematic reviewOral and Maxillofacial Surgery, 2022
- Short versus standard implants at sinus augmented sites: a systematic review and meta-analysisClinical Oral Investigations, 2022
- Biomechanical effects of dental implant diameter, connection type, and bone density on microgap formation and fatigue failure: A finite element analysisComputer Methods and Programs in Biomedicine, 2021
- Peri‐implant diseases and conditions: Consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri‐Implant Diseases and ConditionsJournal of Periodontology, 2018
- Complete mechanical characterization of an external hexagonal implant connection: in vitro study, 3D FEM, and probabilistic fatigueMedical & Biological Engineering & Computing, 2018
- Comparison of external and internal implant-abutment connections for implant supported prostheses. A systematic review and meta-analysisJournal of Dentistry, 2018
- Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic reviewOral and Maxillofacial Surgery, 2015
- Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: An in vitro laboratory and three-dimensional finite element analysisDental Materials, 2012