Microvascular coupler device versus hand‐sewn venous anastomosis: A systematic review of the literature and data meta‐analysis
- 4 April 2020
- journal article
- review article
- Published by Wiley in Microsurgery
- Vol. 40 (5) , 608-617
Abstract
Background: Successful vascular anastomosis is essential for the survival of free tissue transfer. The aim of the study is to review the current literature and perform a meta‐analysis to assess the potential advantages of a mechanical anastomosis coupler device (MACD) over the hand‐sewn (HS) technique for venous anastomoses.Methods: A systematic Medline search was performed to gather all reports of articles related to MACD from 1984 until now. The following data were extracted: first author and publication date, study design, number of patients and anastomosis, coupler size, site and type of reconstruction, venous anastomotic time, flap failure. A meta‐analysis was performed on articles that met the following inclusion criteria: studies comparing MACD and HS technique in venous anastomosis, reporting anastomotic time, and postoperative complications.Results: Thirty‐three studies were included for the analysis. Twenty‐four were retrospective case series and nine were retrospective comparative studies. A total of 12,304 patients were enrolled with a mean age of 49.23 years (range 31–72). A total of 13,669 flaps were accomplished. The thrombosis rate recorded with MACD was 1.47%. The meta‐analysis revealed that MACD significantly decreased anastomotic time (standard difference in means = −0.395 ± 0.105; Z = −3.776; p < .001) and postoperative flap failure risk (odds ratio [OR] = 0.362, 95% confidence interval [CI] = 0.218–0.603, Z = −3.908, p < .001), but it did not decrease postoperative venous thrombosis risk (OR = 0.504, 95% CI = 0.255–1.129, Z = −1.666, p = .096).Conclusions: MACDs are a safe and effective alternative to traditional anastomosis. The anastomotic coupler is easier, much faster, and requires less technical skills than a HS microvascular anastomosis.Keywords
This publication has 49 references indexed in Scilit:
- Venous Anastomoses Using Microvascular Coupler in Free Flap Head and Neck ReconstructionJournal of Oral and Maxillofacial Surgery, 2012
- Recommendations for examining and interpreting funnel plot asymmetry in meta-analyses of randomised controlled trialsBMJ, 2011
- 1000 Consecutive Venous Anastomoses Using the Microvascular Anastomotic Coupler in Breast ReconstructionPlastic and Reconstructive Surgery, 2010
- Use of the Vascular Closure Staple Clip Applier for Microvascular Anastomosis in Free-Flap SurgeryPlastic and Reconstructive Surgery, 2000
- Clinical use of the 3m 2.5 mm mechanical microcoupling device in free tissue transferMicrosurgery, 1994
- Strength of microvascular anastomoses: Comparison between the unilink anastomotic system and suturesMicrosurgery, 1989
- A microvascular anastomotic device: Part II. A Histologic Study In Arteries And VeinsMicrosurgery, 1989
- A microvascular anastomotic device: Part i. a hemodynamic evaluation in rabbit femoral arteries and veinsMicrosurgery, 1989
- A review of vascular anastomosis with mechanical aids and nonsuture techniquesHead & Neck Surgery, 1980
- Experimental nonsuture microvascular anastomosis using a soluble pva tube and plastic adhesiveMicrosurgery, 1979