Enhancements to the Compact Helical System fast ion loss probe
Authors:
- the CHS GroupTGthe CHS Group
- 1 January 1999
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (1) , 838-840
Abstract
A scintillator-based fast ion loss probe has been used to measure 40 keV neutral beam ion loss from Compact Helical System plasmas. Modifications have recently been made to the probe to expand the range of gyroradius covered and to increase the probe acceptance at low pitch angles. In addition, a lamp has been installed inside the probe to facilitate calibration of the scintillator position within the field of view of the video camera. Finally, a Faraday cup structure, integral with the scintillator, has been added to allow direct measurement of the ion current to the probe. This last feature allows much easier absolute calibration of the diagnostic.Keywords
This publication has 9 references indexed in Scilit:
- Alpha particle loss diagnostics in TFTR and tokamak reactorsFusion Engineering and Design, 1997
- Measurement of loss of DT fusion products using scintillator detectors in TFTR (invited)Review of Scientific Instruments, 1995
- Loss of alpha-like MeV fusion products from TFTRNuclear Fusion, 1990
- Measurements of the3He fusion product in TFTRNuclear Fusion, 1989
- Fast ion ripple losses during neutral injection in TFRNuclear Fusion, 1988
- Fusion product measurements of the local ion temperature gradient in the PLT tokamakNuclear Fusion, 1987
- 15 MeV Proton Emission from JET Plasmas Exhibiting Sawtooth Behaviour: Results and InterpretationPhysica Scripta, 1987
- Spatially resolved measurement of alpha particle emission from PLT plasmas heated by ICRHNuclear Fusion, 1985
- Observation of d(d,p)t Reactions in the Princeton Large TorusNuclear Fusion, 1983