Stability of Pedicle Screws After Kyphoplasty Augmentation
Authors:
- Oliver LinhardtOLOliver Linhardt
- Christian LüringCLChristian Lüring
- Jan MatussekJMJan Matussek
- Corinna HambergerCHCorinna Hamberger
- Wolfgang PlitzWPWolfgang Plitz
- Joachim GrifkaJGJoachim Grifka
- 1 April 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Spinal Disorders & Techniques
- Vol. 19 (2) , 87-91
Abstract
The goal of this cadaver study was to compare the stability of pedicle screws after implantation in soft or cured kyphoplasty cement. Pedicle screws were inserted in a total of 30 thoracolumbar vertebrae of 10 different human specimens: 10 screws were implanted in nonaugmented vertebrae (group 1), each 10 screws were placed in soft (group 2) and cured (group 3) cement. Pedicle screws were than evaluated for biomechanical axial pullout resistance. Mean axial pullout strength was 232 N (range 60-600 N) in group 1, 452 N (range 60-1125 N) in group 2 and 367 N (range 112-840 N) in group 3. The paired Student t-test demonstrated a significant difference between pullout strength of groups 1 and 2 (P = 0.0300). Between pullout strength of groups 1 and 3 and between groups 2 and 3 no significant difference was seen. We achieved a 1.9 times higher pullout strength with kyphoplasty augmentation of osteoporotic vertebrae compared with the pullout strength of nonaugmented vertebrae. Implantation of pedicle screws in cured cement is a sufficient method. With this method we found a 1.6 times higher pullout strength then in nonaugmented vertebrae.Keywords
This publication has 25 references indexed in Scilit:
- New Technologies in SpineSpine, 2001
- Revision of Failed Pedicle Screws Using Hydroxyapatite CementSpine, 1998
- Stability of Transpedicle Screwing for the Osteoporotic SpineSpine, 1993
- Complications Associated with the Technique of Pedicle Screw Fixation A Selected Survey of ABS MembersSpine, 1993
- Mechanical Stability of the Pedicle Screw Fixation Systems for the Lumbar SpineSpine, 1992
- An Experimental Study on Transpedicular Screw Fixation in Relation to Osteoporosis of the Lumbar SpineSpine, 1991
- Influence of Bone Mineral Density on the Fixation of Thoracolumbar Implants A Comparative Study of Transpedicular Screws, Laminar Hooks, and Spinous Process WiresSpine, 1990
- Accuracy of Pedicular Screw Placement In VivoSpine, 1990
- Effect of preservation medium on the mechanical properties of cat bonesActa Orthopaedica, 1989
- Roy - Camille Plates in Unstable Spinal ConditionsSpine, 1986