Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs
Authors:
- 1 November 2010
- journal article
- research article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 92-B (11) , 1501-1508
Abstract
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.Keywords
This publication has 50 references indexed in Scilit:
- The painful metal-on-metal hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2009
- The Exeter Universal cemented femoral component at 15 to 17 yearsThe Journal of Bone and Joint Surgery. British volume, 2009
- Metal ion levels after metal-on-metal Ring total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2007
- Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hipThe Journal of Bone and Joint Surgery. British volume, 2007
- Current in vivo wear of metal‐on‐metal bearings assessed by exercise‐related rise in plasma cobalt levelJournal of Orthopaedic Research, 2006
- A benign psoas mass following metal-on-metal resurfacing of the hipThe Journal of Bone and Joint Surgery. British volume, 2006
- Metal ion levels after metal-on-metal proximal femoral replacementsThe Journal of Bone and Joint Surgery. British volume, 2005
- Two distinct cellular mechanisms of osteoclast formation and bone resorption in periprosthetic osteolysisJournal of Orthopaedic Research, 2003
- Wear Is a Function of Use, Not TimeClinical Orthopaedics and Related Research, 2000
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977