Accuracy of Intra-arterial and Subcutaneous Continuous Glucose Monitoring in Postoperative Cardiac Surgery Patients in the ICU
Open Access
- 23 December 2014
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 9 (3) , 663-667
Abstract
Background: The GluCath® intra-arterial continuous glucose monitoring (IA-CGM) system uses a novel quenched chemical fluorescence sensing mechanism to optically measure blood glucose when deployed in the radial artery. The aim of this study was to compare the accuracy of the IA-CGM and the FreeStyle Navigator® subcutaneous continuous glucose monitoring (SC-CGM) system with standard care. Methods: After admission to the intensive care unit (ICU), the IA-CGM was inserted via a 20 gauge radial arterial study catheter and the SC-CGM was placed at the abdominal wall of postoperative cardiac surgery patients with an expected ICU length of stay > 24 hours. Each device was calibrated according to manufacturer instructions. Glucose values of both CGM systems were blinded for the clinical staff. Reference blood glucose samples were collected from the study catheter every 1-2 hours for at least 24 hours and analyzed on a Radiometer ABL blood gas analyzer. Results: The IA-CGM and SC-CGM sensors were successfully inserted in 8 subjects. Accuracy assessment was performed with 183 paired points: 85.8% of the IA-CGM measurements and 84.2% of the SC-CGM measurements met ISO 15197:2003 glucometer criteria (within 20%) across a 79-248 mg/dl (4.4-13.8 mmol/L) glucose range. Overall ± SD mean absolute relative difference was 12.3 ± 11.3% for IA-CGM and 11.1 ± 8.3% for SC-CGM (difference −1.2%, 95% CI −3.3 to 0.8; P = .24). Conclusions: The IA-CGM system directly measured arterial blood glucose and did not interfere with clinical care. However, accuracy was similar to that of the less invasive SC-CGM device.Keywords
This publication has 16 references indexed in Scilit:
- Microcirculation and its relation to continuous subcutaneous glucose sensor accuracy in cardiac surgery patients in the intensive care unitThe Journal of Thoracic and Cardiovascular Surgery, 2013
- Feasibility of fully automated closed-loop glucose control using continuous subcutaneous glucose measurements in critical illness: a randomized controlled trialCritical Care, 2013
- Clinical review: Consensus recommendations on measurement of blood glucose and reporting glycemic control in critically ill adultsCritical Care, 2013
- Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients*Critical Care Medicine, 2011
- Accuracy and Reliability of Continuous Glucose Monitoring in the Intensive Care Unit: A Head-to-Head Comparison of Two Subcutaneous Glucose Sensors in Cardiac Surgery PatientsDiabetes Care, 2011
- Toward Understanding Tight Glycemic Control in the ICUChest, 2010
- Haemodynamic alterations in community-acquired sepsis at early stageCritical Care, 2010
- Real-Time Continuous Glucose Monitoring in Critically Ill PatientsDiabetes Care, 2009
- Pre- and Postoperative Accuracy and Safety of a Real-Time Continuous Glucose Monitoring System in Cardiac Surgical Patients: A Randomized Pilot StudyDiabetes Technology & Therapeutics, 2009
- Use of a Computerized Guideline for Glucose Regulation in the Intensive Care Unit Improved Both Guideline Adherence and Glucose RegulationJournal of the American Medical Informatics Association, 2004