Abstract
Standardization of b values allows improved interstudy comparisons on the diagnostic accuracy of diffusion-weighted MR breast examinations. PurposeTo compare the diagnostic quality of diffusion-weighted (DW) imaging schemes with regard to apparent diffusion coefficient (ADC) accuracy, ADC precision, and DW imaging contrast-to-noise ratio (CNR) for different types of lesions and breast tissue. Materials and MethodsInstitutional review board approval and written, informed consent were obtained. Fifty-one patients with histopathologic correlation or follow-up performed with a 3.0-T MR imager were included in this study. There were 112 regions of interest drawn in 24 malignant, 17 benign, 20 cystic, and 51 normal tissue regions. ADC maps were calculated for combinations of 10 b values (range, 0–1250 sec/mm2). Differences in ADC among tissue types were evaluated. The CNRs of lesions at DW imaging were compared for all b values. A repeated-measures analysis of variance was used to assess lesion differentiation. ResultsADCs calculated from b values of 50 and 850 sec/mm2 were 0.99 × 10−3 mm2/sec ± 0.18 (standard deviation), 1.47 × 10−3 mm2/sec ± 0.21...