Limitations of lesion quality estimated by ablation index: An in vitro study
Authors:
- 25 March 2019
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 30 (6) , 926-933
Abstract
Limitations of the ablation index background: Ablation index (AI) is a novel marker of lesion quality from radiofrequency (RF) catheter ablation. However, AI reliability has not been fully validated by experimental data. The aim of the present study is to validate AI reliability for estimating lesion size using different settings for RF parameters: contact angle, power delivery, and contact force (CF).Methods and results: We evaluated the lesion size in porcine hearts (N = 108) after RF application at three different contact angles to the myocardium: perpendicular (90°), oblique (45°), and parallel (0°). At each angle, RF power at 25, 30, and 35 W was applied at target CF values of 5, 15, and 30 g as measured by the CF sensor to reach target AIs of 300, 400, 500, and 600. AI value was significantly correlated with lesion depth, width, and volume (R = 0.84, 0.82, and 0.87, respectively, all P < 0.001). Lesion depth decreased with smaller contact angles (45° and 0°). Furthermore, high‐power RF energy (35 W) resulted in a significantly smaller lesion volume compared with standard‐power energy (30 W). There were no significant differences in lesion size among CF settings.Conclusions: AI was strongly correlated with lesion depth, width, and volume, but only within a small range of contact angles and RF power delivery settings.Keywords
This publication has 14 references indexed in Scilit:
- Improving procedural and one-year outcome after contact force-guided pulmonary vein isolation: the role of interlesion distance, ablation index, and contact force variability in the ‘CLOSE’-protocolEP Europace, 2018
- Evaluation of a Strategy Aiming to Enclose the Pulmonary Veins With Contiguous and Optimized Radiofrequency Lesions in Paroxysmal Atrial FibrillationJACC: Clinical Electrophysiology, 2018
- Determinants of Acute and Late Pulmonary Vein Reconnection in Contact Force–Guided Pulmonary Vein IsolationCirculation: Arrhythmia and Electrophysiology, 2017
- Effects of Open‐Irrigated Radiofrequency Ablation Catheter Design on Lesion Formation and Complications: In Vitro Comparison of 6 Different DevicesJournal of Cardiovascular Electrophysiology, 2013
- Electrical Reconnection After Pulmonary Vein Isolation Is Contingent on Contact Force During Initial TreatmentCirculation: Arrhythmia and Electrophysiology, 2013
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- RF Catheter Ablation: Lessons on LesionsPacing and Clinical Electrophysiology, 2006
- Virtual Anatomy for Atrial Fibrillation AblationJournal of Cardiovascular Electrophysiology, 2006
- A comparison of open irrigated and non-irrigated tip catheter ablation for pulmonary vein isolationEP Europace, 2004
- Basic Aspects of Radiofrequency Catheter AblationJournal of Cardiovascular Electrophysiology, 1994