Off-Pump Transapical Removal of An Embolized Transcatheter Aortic Valve
- 1 May 2017
- journal article
- case report
- Published by SAGE Publications in Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery
- Vol. 12 (3) , 221-223
Abstract
Transcatheter aortic valve replacement is a less invasive alternative for high-risk patients. However, valve embolization is a rare but dreaded complication. We report the successful off-pump retrieval of an embolized valve after transfemoral transcatheter aortic valve replacement through a left anterior thoracotomy. We maintained the embolized valve on the guidewire and snared it using a transapical approach. We then deployed a valve in an adequate position to ensure hemodynamic stability before transapical removal of the embolized valve. Transapical exteriorization of the femoral guidewire offers additional support, particularly in patients with a horizontal aortic annulus.Keywords
This publication has 5 references indexed in Scilit:
- Review of Reported Causes of Device Embolization Following Trans-Catheter Aortic Valve ImplantationThe American Journal of Cardiology, 2015
- A rare complication with Edwards Sapien: Aortic valve embolization in TAVIVascular, 2014
- Transapical explantation of an embolized transcatheter valveInteractive CardioVascular and Thoracic Surgery, 2011
- Outcome of Patients After Transcatheter Aortic Valve EmbolizationJACC: Cardiovascular Interventions, 2011
- Transcatheter Aortic Valve ImplantationJACC: Cardiovascular Interventions, 2009