High‐resolution fiber‐optic dosimeters for microbeam radiation therapy
- 17 April 2017
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 44 (5) , 1965-1968
Abstract
Purpose: A high resolution, water equivalent, optical and passive x‐ray dosimeter has been constructed using plastic scintillator and optical fiber. This dosimeter has a peak edge‐on spatial resolution of 100 μm in one dimension, with a 10 μm resolution dosimeter under investigation. The dosimeter design has a potential application in synchrotron x‐ray microbeam radiation therapy where a high resolution is vital for accurate dose measurements and quality assurance.Methods: BC‐400 plastic scintillator, of thickness 100 μm, was optically coupled to an optical fiber with core diameter 1 mm. The end was coated in optical paint to improve sensitivity. An identical fiber was made without the scintillator to measure the background Cherenkov radiation induced in the fiber, to allow background signal subtraction. The light captured by the fibers was measured by PMTs. The probe system was exposed to a 6 MV, 10 × 10 cm2 LINAC x‐ray field and the beam profile was measured at 100 cm, as well as the depth dose profile.Results: The measured profiles matched well with ionisation chamber data. Important beam parameters such as penumbra width and percent depth dose at various depths matched the ionisation chamber data, within uncertainty.Conclusions: This work demonstrates that high resolutions can be achieved with a scintillation and optical fiber system. The probe is water‐equivalent, passive, energy independent, radiation hard and inexpensive, making it ideal for further improvements for use with microbeam radiation therapy.Keywords
Funding Information
- UOW's Global Challenges Program
This publication has 12 references indexed in Scilit:
- Fiber‐optic detector for real time dosimetry of a micro‐planar x‐ray beamMedical Physics, 2015
- Establishing the suitability of quantitative optical CT microscopy of PRESAGE® radiochromic dosimeters for the verification of synchrotron microbeam therapyPhysics in Medicine & Biology, 2013
- In vivo two-photon microscopy study of short-term effects of microbeam irradiation on normal mouse brain microvasculatureInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- 1370 POSTER Microbeam Radiation Therapy (MRT) applied to rats' brain tumor: finding the best compromise between normal tissue sparing and tumor curingEuropean Journal of Cancer Supplements, 2005
- WE‐D‐T‐617‐04: Development of a New Fiber Optic Scintillator Dosimeter SystemMedical Physics, 2005
- SIC, an intracerebral radiosensitive probe for in vivo neuropharmacology investigations in small laboratory animals: theoretical considerations and practical characteristicsIEEE Transactions on Nuclear Science, 2000
- Comparative measurements on a series of accelerators by the same vendorMedical Physics, 1999
- Direct reading measurement of absorbed dose with plastic scintillators—The general concept and applications to ophthalmic plaque dosimetryMedical Physics, 1996
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurementsPhysics in Medicine & Biology, 1992