Association left ventricular lead and ventricular arrhythmias after upgrade to cardiac resynchronization therapy in patients with implantable cardioverter defibrillators
Open Access
- 5 May 2019
- journal article
- research article
- Published by Wiley in Clinical Cardiology
- Vol. 42 (7) , 670-677
Abstract
Background: There are some controversial reports related to the pro‐arrhythmic or anti‐arrhythmic potential of cardiac resynchronization therapy (CRT) and little is known about the relationship between ventricular arrhythmia (VA) and left ventricular (LV)‐lead threshold.Hypothesis: Upgrade CRT is anti‐arrhythmic effect of VA with implantable cardioverter‐defibrillator (ICD) patients and has a relationship with the incident of VA and LV‐lead threshold.Methods: Among 384 patients with the implantation of CRT‐defibrillator (CRT‐D), 102 patients underwent an upgrade from ICD to CRT‐D. We divided patients into three groups; anti‐arrhythmic effect after upgrade (n = 22), pro‐arrhythmic effect (n = 14), and unchanging‐VA events (n = 66). The VA event was determined by device reports. We described the electrocardiography parameters, LV‐lead characteristics, and clinical outcomes.Results: Before upgrade, the numbers of VA were 305 episodes and the numbers of ICD therapy were 157 episodes. While after upgrade, the numbers of VA were 193 episodes and the number of ICD therapy were 74 episodes. Ventricular tachycardia cycle length (VT‐CL) after upgrade was significantly slower as compared to those with before upgrade. Pro‐arrhythmic group was significantly higher with delta LV‐lead threshold (after 1 month—baseline) as compared to those with anti‐arrhythmic group (0.74 vs −0.21 V). Furthermore, pro‐arrhythmic group was significantly bigger with delta VT‐CL (after 3 months—before 3 months) as compared to those with anti‐arrhythmic group (P = .03).Conclusions: We described upgrade‐CRT was associated with reduction of VA, ICD therapies and VT‐CL. While 14 patients had a pro‐arrhythmic effect and LV lead threshold might be associated with VA‐incidents.Keywords
This publication has 24 references indexed in Scilit:
- Antiarrhythmic Effect of Reverse Ventricular Remodeling Induced by Cardiac Resynchronization TherapyJACC, 2008
- Potential Proarrhythmic Effects of Biventricular PacingJACC, 2005
- Effect of cardiac resynchronization therapy on the incidence of ventricular arrhythmias in patients with an implantable cardioverter-defibrillatorHeart Rhythm, 2005
- The Effect of Cardiac Resynchronization on Morbidity and Mortality in Heart FailureThe New England journal of medicine, 2005
- Cardiac-Resynchronization Therapy with or without an Implantable Defibrillator in Advanced Chronic Heart FailureThe New England journal of medicine, 2004
- Increase in Ventricular Tachycardia Frequency After Biventricular Implantable Cardioverter Defibrillator UpgradeJournal of Cardiovascular Electrophysiology, 2003
- Combined Cardiac Resynchronization and Implantable Cardioversion Defibrillation in Advanced Chronic Heart FailureJAMA, 2003
- Effect of Epicardial or Biventricular Pacing to Prolong QT Interval and Increase Transmural Dispersion of RepolarizationCirculation, 2003
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionThe New England journal of medicine, 2002