Reliability and validity of the instrumental assessment of implant stability in dry human mandibles
Authors:
- 3 March 2009
- journal article
- research article
- Published by Wiley in Journal of Oral Rehabilitation
- Vol. 36 (4) , 279-283
Abstract
Summary The aim of this study was to determine the intra‐ and interobserver reliability and validity of the instrumental assessment of primary dental implant stability, using resonance frequency analysis (RFA). Sixteen tapered implants and 16 cylindrical implants were installed in eight unfixed dry human mandibles (Cawood classification IV/V). Implant stability quotients (ISQ; the outcome variable of RFA) and peak removal torque were determined. Both the intra‐observer reliability and the interobserver reliability of the RFA measurements were fair‐to‐good, while no significant correlations between the ISQ values and removal torque were found. The removal torque of the cylindrical implants was higher than that of the tapered implants. The smallest detectable difference was almost nine ISQ units. Within the limitations of the present dry cadaver study, it was concluded that (i) primary dental implant stability can be assessed reliably with RFA measurements, (ii) the concurrent validity between RFA measurements and removal torque is poor, (iii) cylindrical implants may be more stable than tapered ones and (iv) two subsequent readings of RFA measurements need to differ at least nine ISQ units before the difference between the two measurements can be considered statistically significant. More research is needed to see whether these conclusions can be extrapolated to the clinical situation, including the assessment of implants during function (secondary stability).Keywords
This publication has 24 references indexed in Scilit:
- Implant stability measurement of delayed and immediately loaded implants during healing.Clinical Oral Implants Research, 2004
- Histology of Retrieved Immediately and Early Loaded Oxidized Implants: Light Microscopic Observations after 5 to 9 Months of Loading in the Posterior MandibleClinical Implant Dentistry and Related Research, 2003
- Immediate Occlusal Loading of Brånemark Implants Applied in Various Jawbone Regions: A Prospective, 1‐Year Clinical StudyClinical Implant Dentistry and Related Research, 2001
- Implant Stability Measurements Using Resonance Frequency Analysis in the Grafted Maxilla: A Cross‐Sectional Pilot StudyClinical Implant Dentistry and Related Research, 1999
- A comparison between cutting torque and resonance frequency measurements of maxillary implantsInternational Journal of Oral and Maxillofacial Surgery, 1999
- Stability measurements of one-stage Brånemark implants during healing in mandiblesInternational Journal of Oral and Maxillofacial Surgery, 1999
- Brånemark Novum®: A New Treatment Concept for Rehabilitation of the Edentulous Mandible. Preliminary Results from a Prospective Clinical Follow‐up StudyClinical Implant Dentistry and Related Research, 1999
- A Review of Nondestructive Test Methods and Their Application to Measure the Stability and Osseointegration of Bone Anchored Endosseous ImplantsCritical Reviews in Biomedical Engineering, 1998
- The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibiaClinical Oral Implants Research, 1997
- A histomorghometric study of screw‐shaped and removal torque titanium implants with three different surface topographiesClinical Oral Implants Research, 1995