Multi-chord fiber-coupled interferometer with a long coherence length laser
- 1 March 2012
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 83 (3) , 033506
Abstract
This paper describes a 561 nm laser heterodyne interferometer that provides time-resolved measurements of line-integrated plasma electron density within the range of 1015−1018 cm−2. Such plasmas are produced by railguns on the plasma liner experiment, which aims to produce μs-, cm-, and Mbar-scale plasmas through the merging of 30 plasma jets in a spherically convergent geometry. A long coherence length, 320 mW laser allows for a strong, sub-fringe phase-shift signal without the need for closely matched probe and reference path lengths. Thus, only one reference path is required for all eight probe paths, and an individual probe chord can be altered without altering the reference or other probe path lengths. Fiber-optic decoupling of the probe chord optics on the vacuum chamber from the rest of the system allows the probe paths to be easily altered to focus on different spatial regions of the plasma. We demonstrate that sub-fringe resolution capability allows the interferometer to operate down to line-integrated densities of the order of 5 × 1015 cm−2.Keywords
All Related Versions
This publication has 5 references indexed in Scilit:
- Interferometer density measurements of a high-velocity plasmoidPhysics of Plasmas, 2010
- A contoured gap coaxial plasma gun with injected plasma armatureReview of Scientific Instruments, 2009
- Heterodyne interferometer with unequal path lengthsReview of Scientific Instruments, 2006
- Phenomenon of interference between two light beams propagating in optical fibers having a large path differenceApplied Optics, 1983
- Principles of Optics. By Max Born and Emil Wolf. Pp. 803 120s. 1959. (Pergamon Press)The Mathematical Gazette, 1961