Experimental validation of numerical study on thermoelectric-based heating in an integrated centrifugal microfluidic platform for polymerase chain reaction amplification
- 1 January 2013
- journal article
- conference paper
- Published by AIP Publishing in Biomicrofluidics
- Vol. 7 (1) , 014106
Abstract
A comprehensive study involving numerical analysis and experimental validation of temperature transients within a microchamber was performed for thermocycling operation in an integrated centrifugal microfluidic platform for polymerase chain reaction (PCR) amplification. Controlled heating and cooling of biological samples are essential processes in many sample preparation and detection steps for micro-total analysis systems. Specifically, the PCR process relies on highly controllable and uniform heating of nucleic acid samples for successful and efficient amplification. In these miniaturized systems, the heating process is often performed more rapidly, making the temperature control more difficult, and adding complexity to the integrated hardware system. To gain further insight into the complex temperature profiles within the PCR microchamber, numerical simulations using computational fluid dynamics and computational heat transfer were performed. The designed integrated centrifugal microfluidics platform utilizes thermoelectrics for ice-valving and thermocycling for PCR amplification. Embedded micro-thermocouples were used to record the static and dynamic thermal responses in the experiments. The data collected was subsequently used for computational validation of the numerical predictions for the system response during thermocycling, and these simulations were found to be in agreement with the experimental data to within ∼97%. When thermal contact resistance values were incorporated in the simulations, the numerical predictions were found to be in agreement with the experimental data to within ∼99.9%. This in-depth numerical modeling and experimental validation of a complex single-sided heating platform provide insights into hardware and system design for multi-layered polymer microfluidic systems. In addition, the biological capability along with the practical feasibility of the integrated system is demonstrated by successfully performing PCR amplification of a Group B Streptococcus gene.Keywords
Funding Information
- Defense Advanced Research Projects Agency (HR001-06-1-0500)
This publication has 31 references indexed in Scilit:
- Rapid multi sample DNA amplification using rotary-linear polymerase chain reaction device (PCRDisc)Biomicrofluidics, 2012
- A simple method of fabricating mask-free microfluidic devices for biological analysisBiomicrofluidics, 2010
- Serial siphon valving for centrifugal microfluidic platformsMicrofluidics and Nanofluidics, 2009
- Microfluidic DNA amplification—A reviewAnalytica Chimica Acta, 2009
- Polydimethylsiloxane microfluidic chip with integrated microheater and thermal sensorBiomicrofluidics, 2009
- Analysis and experiment of capillary valves for microfluidics on a rotating diskMicrofluidics and Nanofluidics, 2007
- Miniaturized PCR chips for nucleic acid amplification and analysis: latest advances and future trendsNucleic Acids Research, 2007
- PCR microfluidic devices for DNA amplificationBiotechnology Advances, 2006
- Towards numerical prototyping of labs-on-chip: modeling for integrated microfluidic devicesMicrofluidics and Nanofluidics, 2005
- Numerical simulations of a low power microchannel thermal cycling reactorInternational Journal of Heat and Mass Transfer, 2002