Proximal outflow graft extrinsic compression of the Heartmate 3 centrifugal flow left ventricular assist device resulting in pump dysfunction
Authors:
- 25 September 2019
- journal article
- case report
- Published by Springer Nature in Journal of Artificial Organs
- Vol. 23 (2) , 180-182
Abstract
We herein report an unusual life-threatening complication with the Hearmate 3 centrifugal ventricular assist device: proximal outflow graft extrinsic compression due to external thrombus leading to pump dysfunction and urgent heart transplantation. Low flow alarms without other abnormalities lead to the diagnosis of outflow occlusion. There was no evidence for a twist of the outflow graft, and no evidence for outflow thrombosis. The location of the outflow occlusion was particularly difficult to diagnose due to metallic artifacts around the pump. This report underscores that this complication should be screened in case of persistent low-flow alarms with the Heartmate 3 centrifugal pump.Keywords
This publication has 6 references indexed in Scilit:
- Surgical treatment of late left ventricular assist device outflow obstruction due to extrinsic compressionThe Journal of Heart and Lung Transplantation, 2019
- Avoidance of Outflow Graft Extrinsic Compression in Polytetrafluoroethylene Reinforcements of the Bend Relief Component.The Journal of Heart and Lung Transplantation, 2019
- A tale of the twist in the outflow graft: An analysis from the MOMENTUM 3 trialThe Journal of Heart and Lung Transplantation, 2018
- Two-Year Outcomes with a Magnetically Levitated Cardiac Pump in Heart FailureThe New England journal of medicine, 2018
- Outflow Graft Occlusion of the HeartMate 3 Left Ventricular Assist DeviceCirculation: Heart Failure, 2017
- A Fully Magnetically Levitated Circulatory Pump for Advanced Heart FailureThe New England journal of medicine, 2017