Evaluating the efficiency of desflurane reflection in two commercially available reflectors
Open Access
- 8 July 2016
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 32 (4) , 605-614
Abstract
With the AnaConDa™ and the MIRUS™ system, volatile anesthetics can be administered for inhalation sedation in intensive care units. Instead of a circle system, both devices use anesthetic reflectors to save on the anesthetic agent. We studied the efficiency of desflurane reflection with both devices using different tidal volumes (VT), respiratory rates (RR), and ‘patient’ concentrations (CPat) in a bench study. A test lung was ventilated with four settings (volume control, RR × VT: 10 × 300 mL, 10 × 500 mL, 20 × 500 mL, 10 × 1000 mL). Two different methods for determination of reflection efficiency were established: First (steady state), a bypass flow carried desflurane into the test lung (flowin), the input concentration (Cin) was varied (1–17 vol%), and the same flow (flowex, Cex) was suctioned from the test lung. After equilibration, CPat was stored online and averaged; efficiency [%] was calculated \((100 \times \left( {1 - \left( {{\text{flow}}_{\text{in}} \times \left( {{\text{C}}_{\text{in}} - {\text{C}}_{\text{ex}} } \right)/{\text{C}}_{\text{Pat}} \times {\text{RR}} \times {\text{V}}_{\text{T}} } \right)} \right)\). Second (washout), flowin and flowex were stopped, the decline of CPat was measured; efficiency was calculated from the decay constant of the exponential regression equation. Both measurement methods yielded similar results (Bland–Altman: bias: −0.9 %, accuracy: ±5.55 %). Efficiencies higher than 80 % (>80 % of molecules exhaled are reflected) could be demonstrated in the clinical range of CPat and VT. Efficiency inversely correlates with the product of CPat and VT which can be imagined as the volume of anesthetic vapor exhaled by the patient in one breath, but not with the respiratory frequency. Efficiency of the AnaConDa™ was higher for each setting compared with the MIRUS™. Desflurane is reflected by both reflectors with efficiencies high enough for clinical use.
Keywords
This publication has 11 references indexed in Scilit:
- Survival after long-term isoflurane sedation as opposed to intravenous sedation in critically ill surgical patientsEuropean Journal of Anaesthesiology, 2016
- Comparing charcoal and zeolite reflection filters for volatile anaestheticsEuropean Journal of Anaesthesiology, 2015
- A novel device for target controlled administration and reflection of desflurane – the Mirus ™Anaesthesia, 2014
- Technical Performance and Reflection Capacity of the Anaesthetic Conserving Device—A Bench Study with Isoflurane and SevofluraneJournal of Clinical Monitoring and Computing, 2009
- The Predictive Performance of a Pharmacokinetic Model for Manually Adjusted Infusion of Liquid Sevofluorane for Use with the Anesthetic-Conserving Device (AnaConDa): A Clinical StudyAnesthesia & Analgesia, 2008
- Inhalational anaesthetics in the ICU: theory and practice of inhalational sedation in the ICU, economics, risk-benefitBest Practice & Research Clinical Anaesthesiology, 2005
- Prolonged isoflurane sedation of intensive care unit patients with the Anesthetic Conserving DeviceCritical Care Medicine, 2004
- Desflurane compared with propofol for postoperative sedation in the intensive care unit † ‡British Journal of Anaesthesia, 2003
- Mathematics and Statistics in Anaesthesia1998
- Elderly patients recover more rapidly from desflurane than from isoflurane anesthesiaJournal of Clinical Anesthesia, 1992