Use of the CEARXRF GUI‐Based Monte Carlo–Library Least‐Squares (MCLLS) Code for the Micro‐Focused EDXRF analyzer
- 5 October 2011
- journal article
- research article
- Published by Wiley in X-Ray Spectrometry
- Vol. 40 (6) , 405-410
Abstract
The CEARXRF GUI‐Based Monte Carlo–Library Least‐Squares (MCLLS) Code is demonstrated with results from a micro‐focused EDXRF analyzer, which can be used to calculate elemental weight fractions in metal alloys or rock samples accurately by library least‐squares regression of the measured X‐ray spectrum with computer‐generated elemental library spectra. An elemental stratified sampling variance reduction technique has been implemented in the CEARXRF5 code, which equalizes the statistical precision of the elemental libraries within the measured sample independent of the relative elemental amounts that are present. Also, an improved Si(Li) detector response function (DRF) has been obtained for micro‐focused X‐ray fluorescence (XRF) analyzers, and the DRF parameters are obtained based on regression from pure elemental experimental spectra. It is demonstrated that the resulting MCLLS approach can greatly improve the accuracy of elemental XRF analysis results. Copyright © 2011 John Wiley & Sons, Ltd.Keywords
This publication has 8 references indexed in Scilit:
- High Definition X‐Ray Fluorescence: Principles and TechniquesX-Ray Optics and Instrumentation, 2008
- Using the Monte Carlo – Library Least-Squares (MCLLS) approach for the in vivo XRF measurement of lead in boneNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- Doubly curved crystal (DCC) X-ray optics and applicationsPowder Diffraction, 2002
- Preliminary studies on combining the K and L XRF methods for in vivo bone lead measurementApplied Radiation and Isotopes, 2001
- A semi-empirical model for the X-ray Si(Li) detector response functionNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1986
- Introduction to X-Ray Spectrometric Analysis1978
- Calculation methods for fluorescent x-ray spectrometry. Empirical coefficients versus fundamental parametersAnalytical Chemistry, 1968
- Some problems in the analysis of steels by X-ray fluorescenceBritish Journal of Applied Physics, 1952