Analysis of factors affecting the accuracy of low-dose insulin dosage using syringes and vials: a cross-sectional study in a Japanese regional hospital
Authors:
- 9 May 2024
- journal article
- research article
- Published by Springer Nature in Diabetology International
- Vol. 15 (3) , 544-549
Abstract
While several studies have shown that insulin pens are more convenient and accurate than conventional administration with syringes and vials (syringes/vials), there is a frequent need for low-dose insulin injections administered by nurses using syringes/vials in hospital settings, particularly for critically ill patients. However, there is a lack of research investigating factors related to the accuracy of low-dose insulin administration using syringes/vials, particularly in hospital settings. We therefore performed a cross-sectional study to assess the accuracy of low-dose insulin administration by registered nurses using syringes/vials and to determine whether time of day, years of experience and adherence to proper injection procedures (vertical insertion/drawing and air bubble checking) affected the accuracy. The participants were 33 registered nurses working in the diabetes ward, and a total of 198 trials were analyzed. Using syringes/vials, the median errors converted to insulin units were found to be 0.6 units for the 2- and 6-unit target doses, and 0.7 units for the 10-unit target dose. In cases with the largest error, errors for the 2-, 6-, and 10-unit target doses were observed to be 2.3, 4.0, and 3.3 units, respectively. In our study, time of day, years of experience and vertical insertion/drawing did not correlate with errors, but errors were significant in the participants who did not check for air bubbles. Nurses can make non-negligible dosage errors when administering low-dose insulin using syringes/vials, and this is particularly likely when air bubble checks are missed.Keywords
This publication has 20 references indexed in Scilit:
- Hospital medication errors: a cross-sectional studyInternational Journal for Quality in Health Care, 2020
- The Hawthorne effect on adherence to hand hygiene in patient careJournal of Hospital Infection, 2020
- Medication Administration Errors in a University HospitalJournal of Patient Safety, 2016
- Investigation of the freely available easy-to-use software ‘EZR’ for medical statisticsBone Marrow Transplantation, 2012
- Exploring risk and ease of use for insulin delivery by nursesJournal of Evaluation in Clinical Practice, 2010
- A Comparison of the Handling and Accuracy of Syringe and Vial Versus Prefilled Insulin Pen (FlexPen)Diabetes Technology & Therapeutics, 2009
- Preference for insulin delivery systems among current insulin users and nonusersClinical Therapeutics, 2004
- A multicenter, randomized, open-label, comparative, two-period crossover trial of preference, efficacy, and safety profiles of a prefilled, disposable pen and conventional vial/syringe for insulin injection in patients with type 1 or 2 diabetes mellitusClinical Therapeutics, 2003
- Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringesArchives of Disease in Childhood, 1998
- Safety, efficacy, acceptability of a pre-filled insulin pen in diabetic patients over 60 years oldDiabetes Research and Clinical Practice, 1995