Active Middle Ear Implant Compared With Open-Fit Hearing Aid in Sloping High-Frequency Sensorineural Hearing Loss
Authors:
- Klaus BoeheimKBKlaus Boeheim
- Stefan-Marcel PokSPStefan-Marcel Pok
- Max SchloegelMSMax Schloegel
- Peter FilzmoserPFPeter Filzmoser
- 1 April 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 31 (3) , 424-429
Abstract
To compare 2 open-ear hearing solutions for sloping high-frequency sensorineural hearing loss: open-fit hearing aid (HA) and active middle ear implant (AMEI). Within-subjects prospective design. Tertiary referral hospital. PATIENTS AND DEVICES: Fourteen patients with sloping, high-frequency sensorineural hearing loss were recruited from 39 patients previously implanted with an AMEI and 10 agreed to participate, ranging in age from 44 to 73 years (mean, 59 yr). Patients were selected because their hearing thresholds (500-3,000 Hz) qualified them for both AMEI and open-fit HA use. All patients received a Vibrant Soundbridge (Vibrant MED-EL) for an average of 25.1 months before data collection and used their AMEI on a daily basis. The open-fit HA used in this study was the Delta 8000 (Oticon). Both study devices have been fit with the specific fitting strategies as recommended by the manufacturer. Sound-field hearing thresholds, Freiburger monosyllabic words in quiet, speech reception thresholds for 50% correct recognition for Freiburger numbers and for Oldenburg sentences in quiet, and speech reception thresholds for 50% correct recognition for Oldenburg sentences in noise. Both HA and AMEI conditions showed significantly better sound-field thresholds and speech recognition on monosyllabic word and sentence tests in quiet and in noise than in the unaided condition. A comparison of aided conditions revealed that, in the AMEI-aided condition, high-frequency audibility and speech discrimination scores in quiet and in noise were significantly better than those in the open-fit HA. Both open-fit HAs and AMEIs provided audiologic benefit to patients with sloping high-frequency sensorineural hearing loss. However, despite overlapping indication criteria for the 2 devices, performance with the AMEI was significantly better for the AMEI than for the open-fit HA.Keywords
This publication has 14 references indexed in Scilit:
- Rehabilitation der HochtoninnenohrschwerhörigkeitHNO, 2006
- Rehabilitation for High-Frequency Sensorineural Hearing Impairment in Adults with the Symphonix Vibrant Soundbridge: A Comparative StudyOtology & Neurotology, 2003
- A Middle Ear Implant, the Symphonix Vibrant Soundbridge: Retrospective Study of the First 125 Patients Implanted in FranceOtology & Neurotology, 2003
- Comparison of Different Vibrant Soundbridge Audioprocessors with Conventional Hearing AidsOtology & Neurotology, 2002
- Phase III Clinical Trial Results with the Vibrant Soundbridge Implantable Middle Ear Hearing Device: A Prospective Controlled Multicenter StudyOtolaryngology–Head and Neck Surgery, 2002
- A Multicenter Study of the Vibrant Soundbridge Middle Ear Implant: Early Clinical Results and ExperienceOtology & Neurotology, 2001
- Vibrant Soundbridge System: Ein neuartiges Hörimplantat für Innenohrschwerhörige - Teil 2: Audiologische Ergebnisse -Laryngo-Rhino-Otologie, 2001
- The otologics MET* *Manufactured by Otologics, LLC., Boulder, Co. ossicular stimulatorOtolaryngologic Clinics of North America, 2001
- Total implantation of the implex TICA hearing amplifier implant for high-frequency sensorineural hearing lossOtolaryngologic Clinics of North America, 2001
- Multicenter audiometric results with the vibrant soundbridge,* *Symphonix Devices, San Jose, CA. a semi-implantable hearing device for sensorineural hearing impairmentOtolaryngologic Clinics of North America, 2001