A new method for measuring aerosol nebulizer output using radioactive tracers
Authors:
- St ChandlerSCSt Chandler
- Ck ConnollyCCCk Connolly
- 1 August 1998
- journal article
- research article
- Published by European Respiratory Society (ERS) in European Respiratory Journal
- Vol. 12 (2) , 467-471
Abstract
Reproducibility and comparability of bronchial challenge tests depends critically on accurate assessment of nebulizer output. Evaporation during nebulization means that simple weighing is inaccurate, overestimating the delivered dose of active ingredient. We wanted to quantify this effect in the context of intermittent nebulization, using a dosimeter as used in bronchial provocation tests. Output of three types of nebulizer, from the MEFAR dosimeter, was measured by radioactive tracer, using a standard solution of technetium-99m-pertechnetate (1.5 kBq x mL(-1)) in 4 mL of normal saline. The aerosol was impacted by suction onto a microfilter, and the radioactivity measured. Nebulizers were weighed before and after nebulization. Ratio of nebulized volume calculated from the radioactivity on the filter, to the total volume loss by weight, was expressed as nebulized ratio. The effect on output of two concentrations of methacholine, two tracers of different weights, and change in temperature, were assessed. Nebulized ratio varied between 44.1-71.6%. Results were more consistent within the same type of nebulizer than between different makes. Neither changes in molar concentration nor molecular weight affected nebulizer output or nebulized ratio. Mean nebulized ratio was 58.5%, showing that calibration by weighing, overestimates the delivered dose by a factor of approximately two. Measuring radioactivity eluted from a microfilter, onto which nebulized output had been impacted proved to be a satisfactory method of calibration.Keywords
This publication has 11 references indexed in Scilit:
- Measurement and prediction of drug solvent losses by evaporation from a jet nebuliserClinical Physics and Physiological Measurement, 1992
- Aerosol Delivery During Mechanical Ventilation: A Predictive In-Vitro Lung ModelJournal of Aerosol Medicine, 1992
- Jet and ultrasonic nebuliser output: use of a new method for direct measurement of aerosol output.Thorax, 1990
- Some factors influencing the efficiency of a jet nebuliser systemClinical Physics and Physiological Measurement, 1990
- Osmolality changes in nebulizer solutionsEuropean Respiratory Journal, 1989
- Room temperature influences output from the Wright jet nebulizerEuropean Respiratory Journal, 1989
- Inaccourate calculation of drug output from nebulisersEuropean Journal of Pediatrics, 1989
- A rapid method for the evaluation of diagnostic radioaerosol delivery systemsEuropean Journal of Nuclear Medicine and Molecular Imaging, 1987
- Factors influencing the size distribution of aerosols from jet nebulisers.Thorax, 1983
- Standardization of inhalation provocation tests: Influence of nebulizer output, particle size, and method of inhalationJournal of Allergy and Clinical Immunology, 1981