New Interpolation Formula for Carbon Resistance Thermometry
Authors:
- 1 November 1968
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 39 (11) , 1656-1657
Abstract
A new interpolation formula for use in carbon resistance thermometry is suggested, and the results of fitting the formula to experimental R vs T data for resistors manufactured by the Speer Company are described. The maximum difference between the calculated and measured temperatures was 1%. The formula was also found to give good agreement with measured temperatures when extrapolated to temperatures somewhat above or below the region covered by the calibration points.Keywords
This publication has 3 references indexed in Scilit:
- CORRESPONDENCE A Simple Formula for use with Carbon Thermometers at Low TemperaturesPhilosophical Magazine, 1958
- The Low Temperature Characteristics of Carbon-Composition ThermometersReview of Scientific Instruments, 1952
- A Sensitive and Reproducible Thermometer in the Range 2° to 20°KPhysical Review B, 1951