Influence of internal standard charge state on the accuracy of mass measurements in orthogonal acceleration time‐of‐flight mass spectrometers
- 19 December 2007
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 22 (2) , 151-155
Abstract
Accuracy of mass measurements performed in orthogonal acceleration time‐of‐flight (oa‐TOF) mass spectrometers highly depends on the quality of the signal and the internal calibration. The use of two reference compounds which bracket the targeted unknown, give rise to ions with sufficient signal‐to‐noise ratio while avoiding detector saturation and produce signals of similar intensity as compared to the target is a common requirement which allow a 5 ppm accuracy on a routine basis. Ion charge state is demonstrated here to be an additional and particularly critical parameter. Using internal references of lower charge state than the target ion systematically yielded overestimated data. Errors measured for quadruply charged molecules were in the range 16–18 ppm when mass calibrants were singly charged ions while accuracy was below 5 ppm when references and target ions were in the same charge state. Magnitude of errors was found to increase with the difference in charge state. This phenomenon arises from the orthogonal acceleration of ions in the TOF analyzer, an interface implemented in all TOF mass spectrometers to accommodate continuous beam ionization sources. Copyright © 2007 John Wiley & Sons, Ltd.Keywords
This publication has 29 references indexed in Scilit:
- Accurate mass measurement for the determination of elemental formula—A tutorialMass Spectrometry Reviews, 2005
- Exact mass measurement on an electrospray ionization time‐of‐flight mass spectrometer: error distribution and selective averagingJournal of Mass Spectrometry, 2003
- An introduction to quadrupole–time‐of‐flight mass spectrometryJournal of Mass Spectrometry, 2001
- Atmospheric pressure ionization time-of-flight mass spectrometry coupled with fast liquid chromatography for quantitation and accurate mass measurement of five pharmaceutical drugs in human plasmaJournal of Mass Spectrometry, 2000
- Peer Reviewed: Orthogonal-Injection TOFMS for Analyzing Biomolecules.Analytical Chemistry, 1999
- A preliminary study using fast gradient liquid chromatography coupled to a quadrupole orthogonal time-of-flight mass spectrometerRapid Communications in Mass Spectrometry, 1999
- Accurate mass determinations for the confirmation and identification of organic microcontaminants in surface water using on-line solid-phase extraction liquid chromatography electrospray orthogonal-acceleration time-of-flight mass spectrometryRapid Communications in Mass Spectrometry, 1999
- The use of liquid chromatography combined with a quadrupole time-of-flight analyser for the identification of trace impurities in drug substanceRapid Communications in Mass Spectrometry, 1997
- High-resolution Mass Spectrometry and Accurate Mass Measurements with Emphasis on the Characterization of Peptides and Proteins by Matrix-assisted Laser Desorption/Ionization Time-of-flight Mass SpectrometryJournal of Mass Spectrometry, 1997
- Important Aspects concerning the Quantification of Biomolecules by Time-of-Flight Secondary-Ion Mass SpectrometryApplied Spectroscopy, 1996