Noncontact Screening Methods for the Detection of Narrow Anterior Chamber Angles
Open Access
- 1 June 2015
- journal article
- glaucoma
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Investigative Ophthalmology & Visual Science
- Vol. 56 (6) , 3929-3935
Abstract
Purpose.: Comparing diagnostic accuracy of biomicroscope techniques (van Herick and Smith's tests, evaluating limbal and central anterior chamber depth, respectively) and advanced imaging (Visante OCT and Pentacam) for detection of gonioscopically narrow anterior chamber angles (ACAs). Methods.: A total of 78 subjects with narrow or open ACAs underwent four index tests, performed on both eyes by examiners masked to other test results. Diagnostic performance was compared with gonioscopy, using International Society of Geographical and Epidemiological Ophthalmology (ISGEO) definition of primary angle closure and a classification based on clinical opinion of occludability. Data were analyzed using both the eye and the individual as unit of analysis. Sensitivity, specificity, and partial area under the receiver operating characteristic curve (AUROC) were generated. Results.: Using the eye as the unit of analysis, the van Herick grading cutoff of 25% or less and ISGEO gonioscopic classification achieved 80% (confidence interval [CI] 65 to 89) sensitivity and 92% specificity (CI 80 to 97) for narrow angle detection, with specificity reaching 97% (CI 87 to 100) for a cutoff of less than or equal to 15%. Notably, with a gonioscopic classification based on clinical opinion of occludability, van Herick (≤25%) together with Smith's test (≤2.50 mm) detected 100% of narrow angle subjects. Of the three Pentacam parameters, anterior chamber volume achieved highest test sensitivity of 85% (CI 70 to 94) using the ISGEO definition. Visante OCT ACA had greatest partial AUROC at 90% specificity, also yielding sensitivity and specificity greater than 85% using the Youden-derived cutoff of less than or equal to 20.7°and ISGEO definition. Conclusions.: Van Herick test and Visante OCT ACA exhibited best discrimination between narrow and open angles both alone, and in combination. Van Herick test affords advantages over Visante OCT, showing potential for identifying individuals who may benefit from further gonioscopic assessment in a case-finding or screening setting.Keywords
This publication has 32 references indexed in Scilit:
- An Analytic Expression for the Binormal Partial Area under the ROC CurveAcademic Radiology, 2012
- Assessment of narrow angles by gonioscopy, Van Herick method and anterior segment optical coherence tomographyJapanese Journal of Ophthalmology, 2011
- Comparison of Slitlamp Optical Coherence Tomography and Scanning Peripheral Anterior Chamber Depth Analyzer to Evaluate Angle Closure in Asian EyesArchives of Ophthalmology (1950), 2009
- Comparison of Gonioscopy and Anterior Segment Ocular Coherence Tomography in Detecting Angle Closure in Different Quadrants of the Anterior Chamber AngleOphthalmology, 2008
- Comparison of the Scanning Peripheral Anterior Chamber Depth Analyzer and the Modified van Herick Grading System in the Assessment of Angle ClosureOphthalmology, 2007
- Detection of Primary Angle Closure Using Anterior Segment Optical Coherence Tomography in Asian EyesOphthalmology, 2007
- Detection of Narrow Angles and Established Angle Closure In Chinese Residents of Singapore: Potential Screening TestsAmerican Journal of Ophthalmology, 2006
- Sample size calculation should be performed for design accuracy in diagnostic test studiesJournal of Clinical Epidemiology, 2005
- Noncontact Goniometry With Optical Coherence TomographyArchives of Ophthalmology (1950), 2005
- Anterior Chamber Depth Measurement as a Screening Tool for Primary Angle-closure Glaucoma in an East Asian PopulationArchives of Ophthalmology (1950), 2000