Abstract
There are increasing concerns about the influence of human bias on the measurement of polyp size at colonoscopy.1–3 A study of 92 124 individual colorectal polyp measurements from the UK bowel cancer screening programme showed that endoscopists, radiologists and pathologists all exhibited terminal digit preference and clustering of polyp measurements at 5mm intervals.1 Although human bias is an acknowledged contributor to the measurement error of polyps, the importance of technology bias remains understated.1–3 Figure 1 illustrates the extent of image distortion from the fish eye lens of modern colonoscopes, where objects in the centre of the display appear magnified and objects at the periphery appear small and warped. It also demonstrates how makeshift measurement tools, such as snare catheters, potentially worsen measurement error from technology bias coupled with illusions of relative size. We hypothesised that both human and technology biases could be addressed by improvements in current colonoscope technology.