Risk Factors for Venous Obstruction in Children with Transvenous Pacing Leads

Authors:

Abstract
To determine tbe incidence and risk factors for venous obstruction in children with transvenous pacing leads, 63 children were evaluated clinically and echocardiographically. Patients with abnormal clinical and/or echocardiographic findings were further investigated by venography. Thirteen patients (21%) had evidence of venous obstruction. Venography in 11 (2 refused) showed that severity of obstruction (as defined by percentage of luminal narrowing) was complete (100%) in 3, severe (> 90%) in 4. and moderate (60%‐90%) in 5 (1 patient having 2 sites of obstruction). Risk factors for obstruction in 55 patients with single implantation procedures (10 with ohstruction: 18%) were sought. Total cross‐sectional area of lead(s) was indexed to body surface area at implantation (INDEX). Patients with obstruction had a higher mean INDEX (7.6 ± 1.6 mm2/m2) than patients without obstruction (4.9 ± 2.0 mm2/m2); P < 0.0002). Receiver‐operator characteristic curves showed an INDEX > 6.6 mm2/m2 to best predict obstruction, with a sensitivity of 90% and specificity of 84%. Since pacingis lifelong, sizing of transvenous leads to the child is important to prevent obstruction and preserve venous access.