The integration of the CIVA simulation platform with the Insightec ExAblate system requires the conversion of data between software systems that can including: skull ge- ometry, material properties, transducer element layout, element phase information, and sonication parameters. This report details the processes and information required to perform simulation of clinical transcranial ultrasound using input from the Insightec Ex- Ablate system and patient data, with the goal of demonstrating and establishing CIVA as a tool for use in clinical use of transcranial ultrasound.Chapters 2 and 3 in this report provide details on the methods by which the skull may be imported and properly positioned within the CIVA environment, as well as methods for the easy import of transducer element phases and amplitude from the Insightec into CIVA Healthcare. The following chapters then present the methods, results, and analysis from using of the CIVA software and import tools in order to simulate clinical scenarios using real human skull samples. The use of imported phase profiles for transcranial focusing is first verified by simulating acoustic steering in water, without the presence of the skull.