MICRA: A Compact Permanent Magnet Fourier Transform Ion Cyclotron Resonance Mass Spectrometer
- 1 April 2004
- journal article
- Published by SAGE Publications in European Journal of Mass Spectrometry
- Vol. 10 (2) , 155-162
Abstract
MICRA,1 a compact Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometer is described. The amount of miniaturisation in this device, based on a 1.24 T permanent magnet, remains compatible with genuine FT-ICR performance and analytical power in the mass range 2–1000 m/z, with a mass resolving power of 73,000 at mass 132. A first application of the transportability is the repetitive coupling of MICRA with a large-scale source of IR photons, the free electron laser CLIO.Keywords
This publication has 15 references indexed in Scilit:
- Permanent magnet applicationsJournal of Magnetism and Magnetic Materials, 2002
- Milestones in fourier transform ion cyclotron resonance mass spectrometry technique developmentInternational Journal of Mass Spectrometry, 2000
- simion for the personal computer in reflectionInternational Journal of Mass Spectrometry, 2000
- Determination of 3,3′-Dichlorobenzidine and its degradation products in environmental samples with a small low-field Fourier transform ion cyclotron resonance mass spectrometerJournal of the American Society for Mass Spectrometry, 1999
- Characterization of long-chain carboxylic esters with CH3OBOCH3+ in a small fourier-transform ion cyclotron resonance mass spectrometerJournal of the American Society for Mass Spectrometry, 1996
- Prediction of a Space Charge Induced Upper Molecular Mass Limit Towards Achieving Unit Mass Resolution in Fourier Transform Ion Cyclotron Resonance Mass SpectrometryJournal of Mass Spectrometry, 1996
- Analysis of space-charge effects in cyclotron resonance mass spectrometry as coupled gyrator phenomenaInternational Journal of Mass Spectrometry and Ion Processes, 1995
- Differentiation of stereoisomeric diols by using CH3OB+OCH3 in a small Fourier transform ion cyclotron resonance mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1995
- Characterization of a small FTICR mass spectrometer based on a permanent magnetAnalytical Chemistry, 1993
- The Measurement ofby Cyclotron ResonancePhysical Review B, 1951