A Guide to Measuring Heart and Respiratory Rates Based on Off-the-Shelf Photoplethysmographic Hardware and Open-Source Software
Open Access
Abstract
The remote monitoring of vital signs via wearable devices holds significant potential for alleviating the strain on hospital resources and elder-care facilities. Among the various techniques available, photoplethysmography stands out as particularly promising for assessing vital signs such as heart rate, respiratory rate, oxygen saturation, and blood pressure. Despite the efficacy of this method, many commercially available wearables, bearing Conformité Européenne marks and the approval of the Food and Drug Administration, are often integrated within proprietary, closed data ecosystems and are very expensive. In an effort to democratize access to affordable wearable devices, our research endeavored to develop an open-source photoplethysmographic sensor utilizing off-the-shelf hardware and open-source software components. The primary aim of this investigation was to ascertain whether the combination of off-the-shelf hardware components and open-source software yielded vital-sign measurements (specifically heart rate and respiratory rate) comparable to those obtained from more expensive, commercially endorsed medical devices. Conducted as a prospective, single-center study, the research involved the assessment of fifteen participants for three minutes in four distinct positions, supine, seated, standing, and walking in place. The sensor consisted of four PulseSensors measuring photoplethysmographic signals with green light in reflection mode. Subsequent signal processing utilized various open-source Python packages. The heart rate assessment involved the comparison of three distinct methodologies, while the respiratory rate analysis entailed the evaluation of fifteen different algorithmic combinations. For one-minute average heart rates’ determination, the Neurokit process pipeline achieved the best results in a seated position with a Spearman’s coefficient of 0.9 and a mean difference of 0.59 BPM. For the respiratory rate, the combined utilization of Neurokit and Charlton algorithms yielded the most favorable outcomes with a Spearman’s coefficient of 0.82 and a mean difference of 1.90 BrPM. This research found that off-the-shelf components are able to produce comparable results for heart and respiratory rates to those of commercial and approved medical wearables.Keywords
This publication has 63 references indexed in Scilit:
- At the Crossroads! Time to Start Taking Smartwatches SeriouslyThe American Journal of Cardiology, 2022
- Wearable smart devices in cancer diagnosis and remote clinical trial monitoring: Transforming the healthcare applicationsDrug Discovery Today, 2022
- Robust PPG motion artifact detection using a 1-D convolution neural networkComputer Methods and Programs in Biomedicine, 2020
- The growing role of Internet of Things in healthcare wearables2020
- Detection and Removal of Motion Artifacts in PPG SignalsMobile Networks and Applications, 2019
- The value of vital sign trends for detecting clinical deterioration on the wardsResuscitation, 2016
- Relationship Between Measurement Site and Motion Artifacts in Wearable Reflected PhotoplethysmographyJournal of Medical Systems, 2010
- Skin Type Classification Systems Old and NewDermatologic Clinics, 2009
- The vexatious vital: Neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triageAnnals of Emergency Medicine, 2005
- Association between clinically abnormal observations and subsequent in-hospital mortality: a prospective studyResuscitation, 2004