Accuracy and precision of calibrated arterial pulse contour analysis in patients with subarachnoid hemorrhage requiring high-dose vasopressor therapy: a prospective observational clinical trial
Authors:
- Sebastian M MetzelderSMSebastian M Metzelder
- Mark CoburnMCMark Coburn
- Christian StoppeCSChristian Stoppe
- Michael FriesMFMichael Fries
- Tim-Philipp SimonTSTim-Philipp Simon
- Marcus Ht ReingesMRMarcus Ht Reinges
- Anke HölligAHAnke Höllig
- Rolf RossaintRRRolf Rossaint
- Gernot MarxGMGernot Marx
- Steffen RexSRSteffen Rex
Open Access
- 5 February 2014
- journal article
- clinical trial
- Published by Springer Nature in Critical Care
- Vol. 18 (1) , R25
Abstract
Introduction: Calibrated arterial pulse contour analysis has become an established method for the continuous monitoring of cardiac output (PCCO). However, data on its validity in hemodynamically instable patients beyond the setting of cardiac surgery are scarce. We performed the present study to assess the validity and precision of PCCO-measurements using the PiCCO™-device compared to transpulmonary thermodilution derived cardiac output (TPCO) as the reference technique in neurosurgical patients requiring high-dose vasopressor-therapy. Methods: A total of 20 patients (16 females and 4 males) were included in this prospective observational clinical trial. All of them suffered from subarachnoid hemorrhage (Hunt&Hess grade I-V) due to rupture of a cerebral arterial aneurysm and underwent high-dose vasopressor therapy for the prevention/treatment of delayed cerebral ischemia (DCI). Simultaneous CO measurements by bolus TPCO and PCCO were obtained at baseline as well as 2 h, 6 h, 12 h, 24 h, 48 h and 72 h after inclusion. Results: PCCO- and TPCO-measurements were obtained at baseline as well as 2 h, 6 h, 12 h, 24 h, 48 h and 72 h after inclusion. Patients received vasoactive support with (mean ± standard deviation, SD) 0.57 ± 0.49 μg · kg-1 · min-1 norepinephrine resulting in a mean arterial pressure of 103 ± 13 mmHg and a systemic vascular resistance of 943 ± 248 dyn · s · cm-5. 136 CO-data pairs were analyzed. TPCO ranged from 5.2 to 14.3 l · min-1 (mean ± SD 8.5 ± 2.0 l · min-1) and PCCO ranged from 5.0 to 14.4 l · min-1 (mean ± SD 8.6 ± 2.0 l · min-1). Bias and limits of agreement (1.96 SD of the bias) were −0.03 ± 0.82 l · min-1 and 1.62 l · min-1, resulting in an overall percentage error of 18.8%. The precision of PCCO-measurements was 17.8%. Insufficient trending ability was indicated by concordance rates of 74% (exclusion zone of 15% (1.29 l · min-1)) and 67% (without exclusion zone), as well as by polar plot analysis. Conclusions: In neurosurgical patients requiring extensive vasoactive support, CO values obtained by calibrated PCCO showed clinically and statistically acceptable agreement with TPCO-measurements, but the results from concordance and polar plot analysis indicate an unreliable trending ability.Keywords
This publication has 39 references indexed in Scilit:
- Guidelines for the Management of Aneurysmal Subarachnoid HemorrhageStroke, 2012
- Clinical validation of a new thermodilution system for the assessment of cardiac output and volumetric parametersCritical Care, 2012
- Effect of norepinephrine dosage and calibration frequency on accuracy of pulse contour-derived cardiac outputCritical Care, 2011
- A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac OutputAnesthesia & Analgesia, 2010
- Cardiac output derived from arterial pressure waveformCurrent Opinion in Anaesthesiology, 2009
- Self-calibrating pulse contour cardiac output: do validation studies really show its clinical reliability?Critical Care, 2009
- Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac outputCritical Care, 2009
- Validation of continuous cardiac output technologies: consensus still awaitedCritical Care, 2009
- The Effects of Vasodilation on Cardiac Output Measured by PiCCOJournal of Cardiothoracic and Vascular Anesthesia, 2008
- An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgeryAnaesthesia, 2007