Measurements of wavelength shifting (WLS) fibre readout for a highly multiplexed, depth-encoding PET detector
Authors:
- Huini DuHDHuini Du
- Yongfeng YangYYYongfeng Yang
- Simon R CherrySCSimon R Cherry
- 10 April 2007
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (9) , 2499-2514
Abstract
We are attempting to develop a laboratory PET scanner for mouse imaging that utilizes far fewer detectors and channels of electronics, thus reducing cost, whilst retaining state-of-the-art performance. The detectors are based on LSO arrays read out by wavelength shifting (WLS) fibres placed on the top and the bottom of the arrays. Depth of interaction information will be obtained from the ratio of the signals at either end of the array. For acceptable performance, it is critical to maximize collection of light photons from the ends of the fibres and to minimize the optical crosstalk between adjacent fibres. Factors which can affect the light collection and crosstalk were studied, including coupling materials between fibres and crystals, reflectors wrapped around the fibre sides and ends, fibre size and shape, and number of layers of fibre cladding. Properties of WLS fibres such as the transmission attenuation and transverse absorption were also studied. The light yield from 2 x 2 x 10 mm(3) LSO crystals collected from the end of a 2 x 2 x 30 mm(3) WLS fibre was up to 24% (typical values 16-20%) of that obtained by direct coupling of the LSO crystal. This light collection efficiency appears to be sufficient for decoding interaction position in these detectors.Keywords
This publication has 13 references indexed in Scilit:
- Evaluation of high performance data acquisition boards for simultaneous sampling of fast signals from PET detectorsPhysics in Medicine & Biology, 2004
- Performance evaluation of the 16-module quad-HIDAC small animal PET cameraPhysics in Medicine & Biology, 2004
- Small Animal PET Systems2004
- Evaluation of a position sensitive avalanche photodiode for PETIEEE Transactions on Nuclear Science, 2003
- MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imagingPhysics in Medicine & Biology, 2003
- The Rise of the Mouse, Biomedicine's Model MammalScience, 2000
- Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolutionPhysics in Medicine & Biology, 1999
- MicroPET: a high resolution PET scanner for imaging small animalsIEEE Transactions on Nuclear Science, 1997
- Simple charge division readouts for imaging scintillator arrays using a multi-channel PMTIEEE Transactions on Nuclear Science, 1996
- Design studies for a PET detector module using a PIN photodiode to measure depth of interactionIEEE Transactions on Nuclear Science, 1994