Mechatronic Design and Implementation of a Low-Noise Powered Knee Prosthesis With High Backdrivability
- 4 April 2023
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE/ASME Transactions on Mechatronics
- Vol. 28 (6) , 3180-3190
Abstract
Working noise is an important consideration that hinders the acceptance of powered prostheses by amputees. In this article, we present a powered knee prosthesis with low noise and high backdrivability. The powertrain of the proposed prosthesis consists of a motor and a four-stage synchronous belt to decrease working noise. At the same time, benefiting from the low impedance of the motor and low impedance design, the proposed prosthesis is endowed with the characteristics of low impedance, and thus, high backdrivability. A three-layer control method based on a finite state machine is implemented, and walking experiments have been carried out on a transfemoral amputee. The results show the characteristics of the proposed prosthesis in low impedance, low noise, and the ability to reproduce a natural gait. In particular, the working noise of the proposed prosthesis is less than 46 dBA most time, which is lower than common household appliances, and may have positive effects on the acceptance of daily uses.Keywords
Funding Information
- National Natural Science Foundation of China (51922015, 91948302, 91648207)
- National Key R&D Program of China (2018YFB1307300)
This publication has 36 references indexed in Scilit:
- Virtual Constraint Control of a Powered Prosthetic Leg: From Simulation to Experiments With Transfemoral AmputeesIEEE Transactions on Robotics, 2014
- Clutchable series-elastic actuator: Implications for prosthetic knee designThe International Journal of Robotics Research, 2014
- Running With a Powered Knee and Ankle ProsthesisIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2014
- Design and Control of a Pneumatic Artificial Muscle Actuated Above-Knee ProsthesisJournal of Medical Devices, 2011
- Evaluation of gait symmetry after stroke: A comparison of current methods and recommendations for standardizationGait & Posture, 2010
- Gear impacts and idle gear noise: Experimental study and non-linear dynamic modelMechanical Systems and Signal Processing, 2009
- Design and Control of a Powered Transfemoral ProsthesisThe International Journal of Robotics Research, 2008
- Studies on the Sound and Vibration of a Ball Screw. (5th Report. Sound Caused by Random Waviness on Flank Surfaces).TRANSACTIONS OF THE JAPAN SOCIETY OF MECHANICAL ENGINEERS Series C, 1995
- Osteoarthritis and elderly amputee gaitArchives of Physical Medicine and Rehabilitation, 1994
- Biomechanics of below-knee amputee gaitJournal of Biomechanics, 1988