In-Source Decay during Matrix-Assisted Laser Desorption/Ionization Combined with the Collisional Process in an FTICR Mass Spectrometer
- 23 July 2013
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 85 (16) , 7809-7817
Abstract
The type of ions detected after in-source decay (ISD) in a MALDI source differs according to the ion source pressure and on the mass analyzer used. We present the mechanism leading to the final ISD ions for a Fourier transform-ion cyclotron resonance mass spectrometer (FTICR MS). The MALDI ion source was operated at intermediate pressure to cool the resulting ions and increase their lifetime during the long residence times in the FTICR ion optics. This condition produces not only c′, z′, and w fragments, but also a, y′, and d fragments. In particular, d ions help to identify isobaric amino acid residues present near the N-terminal amino acid. Desorbed ions collide with background gas during desorption, leading to proton mobilization from Arg residues to a less favored protonation site. As a result, in the case of ISD with MALDI FTICR, the influence of the Arg residue in ISD fragmentation is less straightforward than for TOF MS and the sequence coverage is thus improved. MALDI-ISD combined with FTICR MS appears to be a useful method for sequencing of peptides and proteins including discrimination of isobaric amino acid residues and site determination of phosphorylation. Additionally we also used new software for in silico elimination of MALDI matrix peaks from MALDI-ISD FTICR mass spectra. The combination of high resolving power of an FTICR analyzer and matrix subtraction software helps to interpret the low m/z region of MALDI-ISD spectra. Finally, several of these developed methods are applied in unison toward a MALDI ISD FTICR imaging experiment on mouse brain to achieve better results.Keywords
This publication has 42 references indexed in Scilit:
- Peptide backbone fragmentation initiated by side‐chain loss at cysteine residue in matrix‐assisted laser desorption/ionization in‐source decay mass spectrometryJournal of Mass Spectrometry, 2013
- Selected Protein Monitoring in Histological Sections by Targeted MALDI-FTICR In-Source Decay ImagingAnalytical Chemistry, 2013
- An Analytical Pipeline for MALDI In-Source Decay Mass Spectrometry ImagingAnalytical Chemistry, 2011
- Matrix Sublimation/Recrystallization for Imaging Proteins by Mass Spectrometry at High Spatial ResolutionAnalytical Chemistry, 2011
- Discrimination of Isobaric Leucine and Isoleucine Residues and Analysis of Post-Translational Modifications in Peptides by MALDI In-Source Decay Mass Spectrometry Combined with Collisional CoolingAnalytical Chemistry, 2010
- Ammonium Sulfate and MALDI In-Source Decay: A Winning Combination for Sequencing PeptidesAnalytical Chemistry, 2009
- New Method for Characterizing Highly Disulfide-Bridged Peptides in Complex Mixtures: Application to Toxin Identification from Crude VenomsJournal of Proteome Research, 2007
- Locating and identifying posttranslational modifications by in‐source decay during MALDI‐TOF mass spectrometryProtein Science, 1999
- Fourier transform ion cyclotron resonance mass spectrometry: A primerMass Spectrometry Reviews, 1998
- Ab Initio Studies of Neutral and Protonated Triglycines: Comparison of Calculated and Experimental Gas-Phase BasicityJournal of the American Chemical Society, 1994