A Steerable Needle Technology Using Curved Concentric Tubes
- 1 October 2006
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21530858,p. 2850-2856
Abstract
A new approach to steerable needle design is proposed for use in minimally invasive surgery. The technology is based on sets of curved concentric tubes. By rotating and extending the tubes with respect to each other, the position and orientation of the needle tip, as well as the shape of the inserted length, can be controlled. A mechanics model is presented for computing the shape of the needle. Forward and inverse kinematic equations are also derived. In addition, experimental results are presented as validation of the approachKeywords
This publication has 4 references indexed in Scilit:
- Planning for Steerable Bevel-tip Needle Insertion Through 2D Soft Tissue with Obstacles2006
- Steering flexible needles under Markov motion uncertainty2005
- Various Approaches for CT-guided Percutaneous Biopsy of Deep Pelvic Lesions: Anatomic and Technical ConsiderationsRadioGraphics, 2004
- Needle insertion modeling and simulationIEEE Transactions on Robotics and Automation, 2003