Abstract
Custom polydimethylsiloxane (PDMS) microfluidic devices allow for small-volume human blood research under hemodynamic conditions of bleeding and clotting. However, issues of PDMS molding/assembly, bio-coating, and sample preparation often limit their point-of-care use. We aim to develop a microfluidic device that has the same utility as previously established PDMS devices but which is more usable in point-of-care operation. We designed an injection-molded 1 × 3 in.2 device with eight flow paths crossing a bio-printed surface of a collagen/tissue factor. The device is rapidly primed and compatible with multi-channel pipetting (−1, the intrachip CV at 400 s for platelet and fibrin deposition was 10% and the interdonor CV at 400 s was 30% for platelet and 22% for fibrin deposition (across 10 healthy donors). No significant difference was detected for samples tested on a new chip vs a chip stored for 6 months at 4 °C. Using the fibrin signal, dose–response testing of whole blood revealed IC50's of 120 nM for rivaroxaban and apixaban, and 60 nM for dabigatran. A complete reversal of apixaban inhibition was observed for an equimolar addition of Xa DOAC reversal agent Andexanet Alfa. We demonstrate the ability to manufacture single-use, storage-stable eight-channel chips. In clinical settings, such chips may help evaluate patient bleeding risk, therapy choice, drug activity, or reversal.